16
4)
Press the
∆∆∆∆
or
∇
∇
∇
∇
keys
to scroll the lower display to match the correct
value. The upper display will now show the offset value. You can offset
temperature up to
±
5
o
C.
5)
Press the ENTER key
to confirm your selection.
6) Continue with additional Advanced Set-up procedures, or return to the
Measurement mode by pressing the
∆∆∆∆
and
∇
∇
∇
∇
keys (escape) simultaneously.
7.2.3 Setting manual temperature compensation
Note: This option is not available when the controller is set for ATC
operation.
For manual temperature compensation, two
different temperatures: process and
calibration, can be input independently. This
allows calibration at a temperature other than
the process temperature. Example: setting a
calibration temperature of 25
o
C allows
calibration using standard buffer solutions at
25
o
C, even if the process temperature is
different.
1) Select “ATC oFF” as described above in
section 7.2.1
2)
Press the ENTER key
. The upper display
shows the current process temperature and the lower display shows “P.
o
C”
to indicate process temperature.
3)
Press the
∆∆∆∆
or
∇
∇
∇
∇
ke
ys to adjust the process temperature value, between
–9.9 and 125
o
C.
4) After setting the process temperature value,
press the ENTER key
. The
upper display shows the current calibration temperature and the lower
display shows “C.
o
C”.
25.0
25.0
25.0
25.0
C.C
SETUP
HOLD
oFF
oFF
oFF
oFF
Atc
SETUP
HOLD
25.0
25.0
25.0
25.0
P.C
SETUP
HOLD
Содержание ALPHA PH 800 PHORP CONTROLLERTRANSMITTER
Страница 1: ...EUTECH INSTRUMENTS lpha pH800 pH and ORP Controller Transmitter Operating Instructions 68X216813 03 99...
Страница 4: ......
Страница 11: ...7 0V Pt100 RELAY A L N PE 100mA F FUSE 250VAC HOLD RELAY B Pt1000 PAL S NC NC NC NC...
Страница 34: ...CALIBRATION RECORD Calibration Date Electrode Slope Electrode Offset...