15
7.2 Setting temperature (Set
o
C) sub-function
7.2.1 Selecting automatic or manual temperature
compensation
1)
Enter Advanced Set-up mode
by pressing the ENTER key.
2)
Press the
∆∆∆∆
or
∇
∇
∇
∇
keys
to scroll until the
display shows “Set
o
C”.
3)
Press the ENTER ke
y. The lower display
shows “Atc”; the upper display shows “on” or
“oFF” depending on whether or not ATC is
selected.
4)
Press the
∆∆∆∆
or
∇
∇
∇
∇
keys
to toggle between
ATC on and off.
5) Proceed with additional Advanced Set-up
procedures (press ENTER again) or return to
Measurement mode by pressing the
∆∆∆∆
and
∇
∇
∇
∇
keys (escape) simultaneously.
7.2.2 Temperature calibration (ATC mode only)
1) Select “ATC on” as described above in
Section 7.2.1.
2)
Press the ENTER key
. The upper display
indicates the current temperature offset. The
current measured temperature is shown in the
lower display.
3) Compare the current measured temperature
on the controller display to a thermometer
known to be accurate. Note down the correct
temperature value.
0.0
0.0
0.0
0.0
25.0
ATC
SETUP
HOLD
o
C
on
on
on
on
Atc
SETUP
HOLD
oFF
oFF
oFF
oFF
Atc
on
on
on
on
Atc
SETUP
HOLD
SETUP
SEt
SEt
SEt
SEt
oooo
CCCC
SETUP
HOLD
HOLD
Содержание ALPHA PH 800 PHORP CONTROLLERTRANSMITTER
Страница 1: ...EUTECH INSTRUMENTS lpha pH800 pH and ORP Controller Transmitter Operating Instructions 68X216813 03 99...
Страница 4: ......
Страница 11: ...7 0V Pt100 RELAY A L N PE 100mA F FUSE 250VAC HOLD RELAY B Pt1000 PAL S NC NC NC NC...
Страница 34: ...CALIBRATION RECORD Calibration Date Electrode Slope Electrode Offset...