Instruction Manual MM850470 ed.1a
24
•
Press the [
ª
] key to highlight the “
RTD thermometer
” parameter.
•
Press the [
¨
] key to select the option. Press the [
MENU
] key to return in measure mode without changes.
RTD thermometer
3 wire
4 wire
Press <
¨
> to SET
•
Press the [
ª
] key to highlight the connection for your RTD.
IMPORTANT
P
LEASE SELECT
4
WIRE CONNECTION IF YOU HAVE A
2
WIRE
RTD
AND USE THE ELECTRICAL CONNECTION AS SHOWN ON
2.4
CHAPTER
.
•
Press the [
¨
] key to confirm the selection. Press the [
MENU
] key to return in measure mode without changes.
3 wire
Pt100 IEC
Pt100 OIML
Pt100 USLAB
Pt100 US
Pt100 SAMA
Press <
¨
> to SET
4 wire
Pt100 IEC
Pt100 OIML
Pt100 USLAB
Pt100 US
Pt100 SAMA
Press <
¨
> to SET
•
Press the [
ª
] key to highlight the requested type of RTD.
•
Press the [
¨
] key to confirm the selection and going in measuring mode. Press the [
MENU
] key to return in measure mode without changes.
4.1.7
External Cold Joint Reference Setting
•
Press the [
MENU
] key. The following page is shown: