Instruction Manual MM850470 ed.1a
21
SELECT Ch
In
Out
Press <
¨
> to SET
•
Press the [
ª
] key to highlight the “
In
” parameter.
•
Press the [
¨
] key to select the option. Press the [
MENU
] key to return in measure mode without changes.
SELECT Ch
RTD thermometer
Current
Voltage
Resistance
Press <
¨
> to SET
•
Press the [
ª
] key to highlight the “
Current
” parameter.
•
Press the [
¨
] key to select the option. Press the [
MENU
] key to return in measure mode without changes.
Current
Active loop
Passive loop
Press <
¨
> to SET
•
Press the [
ª
] key to choose the loop mode.
Active current loop
: select this option if you are connecting to an active current loop (your circuit provide to supply the loop)
Passive current loop
: select this option if you are connecting to a passive current loop (the calibrator should provide to supply the loop).