User manual TOM600
Item partnumber : 61-200-0600
4
* mathematical representation of technical representation < 2T
Ω
* * Alternating to the timer display T=OFF is shown the measure voltage mode: Auto, 100V or 10V
To turn off the appliance press again key B .
In the battery mode the unit turns off , if no button was pressed, automatically after approximately
5 minutes.
2.2. Charging the NiMH Battery
The EUROSTAT TOM600 allowed only to operating directly with the enclosed power supply. Also only
with these the installed NiMH – batteries can be charged. Connect the power supply to the integrated socket,
and connect the plug power supply to a power socket. If the appliance is switched off, and the charging work
correct, the
green Charge LED shines.
The NiMH - batteries are completly charged after maximum of 14 hours and good for > 10 hours
continuous operation. A longer charging time can dammage the batteries.
2.3.
Measure range and measure voltage
The EUROSTAT TOM600 has an automatic ranging approve the whole resistor range. The appliance,
accordingly the measure resistance, selects the Measure voltage automatically.
•
Measured resistance less then 200 k
Ω
—-
Measure Voltage 10 VDC
•
Measured resistance more then 200 k
Ω
—-
Measure Voltage 100 VDC
In the Set-Up this automation can be switched off, and a constant measure voltage can be select.
The measure ranges are than as following :
•
Measure voltage 10V -- measure range 20k
Ω
.... 200G
Ω
•
Measure voltage 100V -- measure range 100k
Ω
.... 2T
Ω
3. „Set-up“
Function
Choice of the measure time
In all cases, the tester measures each 0.1 seconds : if the selected time is 3 secondes, 30 measures will be
taken.
Mode “LAPSE” : the last measure is displayed
Mode “AVERAGE” : the mean of all measures is displayed
By pressing the key (A) Set-Up longer as 2 seconds the „set-up" function is starting.
At the display appears:
SET
TIMER !
TIMER OFF
(On)
By pressing key (B) „Change” the timer is switches between off and on.
If pressing key (A) „Set-Up“, by TIMER OFF the timer is switched off and you can continued by point 3.1.
If you press the key by TIMER ON, you can now choose the measure time.
In the display appears :
NEW TIMER !
TIME
=
001
s