Reader check 

The reader check must be performed on the instrument straight away, to validate the 
correct position of the  reader’s  internal components. After the first verification, it is 
recommended to repeat the reader check yearly. 
To do so: 

Insert the chart supplied with the instrument in the QR code adapter (fig. 2) 

Insert the adapter in the adapter inlet on the front 

of the reader as shown as below

 

From the main menu, tap on “Setting”, “Advanced” and then on “Reader Check” 
(fig. 3) 

The message “Please Put Calibration Target” will be shown.  

Select “OK”. 

A screen will appear, like in fig. 4. Please make sure the QR code is correctly 
placed on the left of the screen and the view area (framed in green) on the 
right.  

Tap “Yes”. 

Once the check is completed, two messages can be displayed: 

o

 

“Check Result: Pass”: The camera is correctly positioned; no further 
actions are needed. 

o

 

“Check Result: Fail”: Make sure the reader check chart is placed in 
the right position and that no dust or fingerprints are present on the 
surface. Perform the reader check a second time. If the check still 
fails, please contact Eurofins Technologies local technical assistance. 

 
 

 