U2836 Operation Manual                            Chapter 3 Introduction to [MEAS DISP] 

 12

3.1.2

 

Test range 

Measurement range should be selected in accordance with the impedance value 
of the tested LCR component. 
U2836 has 5 AC measurement ranges: 30

Ω

, 100

Ω

, 1k

Ω

, 10k

Ω

 and 100k

Ω

When the output impedance is selected as 30

Ω

 and 50

Ω

, U2836 has 6 AC 

measurement ranges: 10

Ω

, 30

Ω

, 100

Ω

, 1k

Ω

, 10k

Ω

 and 100k

Ω

Operation steps for setting test range: 

1)

 

Move the cursor to the range zone, the following soft keys will be displayed: 

„

 

AUTO The soft key is used to set the range mode to AUTO. 

„

 

HOLD The soft key is used to switch the AUTO mode to the HOLD mode. 
When the range mode is set to HOLD, the range will be locked in the 
current measurement range. The current measurement range will be 
displayed in the range zone. 

„

 

INCR

+ The soft key is used to increase the range under HOLD mode. 

„

 

DECR

-      The soft key is used to decrease the range under HOLD mode. 

2)

 

Use soft keys to set measurement range. 

3.1.3

 

Test frequency 

The measurement range of U2836 ranges from 50Hz to 200kHz, total 37 
frequency points. 
 

Operation steps for setting test frequency: 

U2836 provides two methods to set measurement frequency. The first one is to 
use soft keys and the other one is to input data by using numeric keys. 
1)

 

Move the cursor to the 

FREQ

 zone, the following soft keys will be displayed. 

„

 

INCR

++ 

This is a coarse adjustment soft key used to increase the frequency. Press 
this key, the frequency will change as the following sequence: 50Hz, 
100Hz, 1kHz, 10kHz, 100kHz, 200kHz. 

„

 

INCR

This is a fine adjustment soft key used to increase the frequency. Press 
this key, the frequency will switch between the following ones: 
50 Hz  100 Hz  

1 kHz   

10 kHz      100 kHz     

60Hz  120 Hz  

1.2 kHz  

12 kHz   

120 kHz 

80 Hz  150 Hz  

1.5 kHz  

15 kHz   

150 kHz 

  200 

Hz 

 2 

kHz 

 20 

kHz 

 200 

kHz 

  250 

Hz 

 2.5 

kHz 

 25 

kHz 

 

  300 

Hz 

 3 

kHz 

 30 

kHz 

 

  400 

Hz 

 4 

kHz 

 40 

kHz 

 

500 Hz  

5 kHz   

50 kHz   

600 Hz  

6 kHz   

60 kHz   

Содержание U2836

Страница 1: ...d d No 1 Ziyang Road Tianning District Changzhou Jiangsu T T Te e el ll 0 0 05 5 51 1 19 9 9 8 8 85 5 55 5 50 0 05 5 51 1 19 9 99 9 9 F F Fa a ax x x 0 0 05 5 51 1 19 9 9 8 8 85 5 55 5 50 0 05 5 51 1...

Страница 2: ...1 MEAS DISP 8 2 4 2 MEAS SETUP 8 2 4 3 SYSTEM 8 2 5 Basic Operation 8 2 6 Start the instrument 9 Chapter 3 Introduction to MEAS DISP 10 3 1 MEAS DISPLAY 10 3 1 1 Test function 11 3 1 2 Test range 12 3...

Страница 3: ...INFORMATION 33 4 1 12 FIRMWARE UPDATE 34 4 1 13 SYSTEM TEST 34 4 2 FILE 35 4 2 1 Setup file for single group component EST 35 4 2 2 U disk manage performance 36 4 2 3 Operation steps for file managem...

Страница 4: ...ation resistance 50 6 4 2 Insulation intensity 51 6 4 3 Leakage current 51 6 5 Electromagnetic compatibility 51 6 6 Performance test 51 6 6 1 Working condition 51 6 6 2 The used instruments and device...

Страница 5: ...panel 2 Power supplying frequency range 47 63Hz 3 Power supplying power range less than 20VA 4 Power supplying input phase line L zero line N ground lead E should be as same as the power plug of the i...

Страница 6: ...5 and 40 humidity is less than 85 RH If the instrument will not be put in use for a time please have it properly packed with its original box or a similar box for storing 6 The instrument especially t...

Страница 7: ...U2836 Operation Manual Chapter 1 Out of Box Audit 3 1 7 Other features 1 Power consumption power 20VA 2 Dimension W H D 280mm 88mm 320mm 3 Weight About 3 5kg...

Страница 8: ...o save or use a specified file 4 MEAS SETUP Press this key to enter into the setup page of measurement setup 5 SYSTEM Press this key to enter into the system setup page 6 Ground terminal The ground te...

Страница 9: ...ic of the input value 14 Numerical keys These keys are used to input data to the instrument The key consists of numerical keys 0 to 9 decimal point and key 15 FAIL indicator LED indicator shows the te...

Страница 10: ...ing used to install power fuse protect instrument Warning Be sure that the direction of fuse is accordant with power supply voltage range before charging 5 Ground terminal The ground terminal is conne...

Страница 11: ...esponding specific tools 3 System icon zone In this zone display the system status U disk is connected Romot control status Keyboard locked Data recording status 4 Soft keys The zone is used to displa...

Страница 12: ...to enter into the system setup page The following soft keys will be available SYSTEM SETUP SYSTEM INFO FIRMWARE UPDATE SYSTEM TEST FILE LIST 2 5 Basic Operation Basic operation of U2836 is as follows...

Страница 13: ...supply voltage and frequency conform to above specifications Power input phase line L zero line N ground line E should be the same as that of the instrument Press the power switch at the left corner o...

Страница 14: ...arameters can be set in this page Test function FUNC Test frequency FREQ Test level LEVEL Test range RANGE Test speed SPEED Output impedance RSOU There are 6 zones in this page FUN FREQ LEVEL RANG SPE...

Страница 15: ...parameters are respectively displayed in two lines in the form of upper case characters The primary parameter displays in the upper line while the secondary parameter displays in the lower line Opera...

Страница 16: ...he soft key is used to decrease the range under HOLD mode 2 Use soft keys to set measurement range 3 1 3Test frequency The measurement range of U2836 ranges from 50Hz to 200kHz total 37 frequency poin...

Страница 17: ...e is to use soft keys while the second one is to input data by numeric keys 1 Move the cursor to LEVEL the following soft keys will be displayed INCR This soft key is used to increase the level of tes...

Страница 18: ...ursor to TOOL zone the following soft keys will be displayed D P FIX A D P FIX B DISP OFF ON EASY START SAVE CSV 2 Press D P FIX A to lock the decimal location of primary parameter TEST result When th...

Страница 19: ...alue HIGH LOW 3 2 1Comparator function U2836 has an inserted compare function which can divide DUT to up to 4 bins from BIN1 to BIN3 and BIN OUT Users can set 3 pairs of primary parameter limit and on...

Страница 20: ...er will be displayed in the help zone Then the following soft keys will be displayed YES NO 5 Press the soft key YES to reset all bin counts to 0 6 Press the soft key NO to cancel the reset operation...

Страница 21: ...r Mode TRIG Delay Time DELAY Average times AVG Auto LCR Function AUTO LCR Voltage Level Monitor ON OFF MON 1 Current Level Monitor ON OFF MON 2 Deviation Test Mode A DEV A Deviation Test Reference Val...

Страница 22: ...f testing when U2836 receives a trigger signal it will be ignored So the trigger signal should be sent after the test is done When optional HANDLER interface triggers U2836 the trigger mode is set as...

Страница 23: ...the correction is switched between OPEN or SHOR the level monitor value will be influenced as well Operation steps for setting the monitor function Execute the following operation steps to set the lev...

Страница 24: ...test terminal has a reliable contact Operation steps for setting the delay function Execute the following steps to set the measurement delay time 1 Move the cursor to the DELAY zone 3 the following so...

Страница 25: ...percentage of the difference between the test value of DUT and the preset reference value divided by the reference value Its calculating formula is as below X Y Y 100 Where X is the test value of DUT...

Страница 26: ...f the secondary parameter 3 4 CORRECTION Open and short correction can be used to eliminate the distribution capacitance spurious impedance and other measurement errors U2836 provides two correction m...

Страница 27: ...inductance under 37 frequencies It will take about 20 seconds to finish the open full frequency correction In the process of correction the following soft key will be displayed CANCEL This soft key ca...

Страница 28: ...ction on the current frequency 5 Press the SWEEP SHORT soft key U2836 will test the short spurious impedances resistance and reactance of 37 frequencies Short full frequency correction takes about 20...

Страница 29: ...parameter FUNC Limit mode of compare function MODE Nominal value NOM Auxiliary bin ON OFF AUX Compare function ON OFF COMP Low limit of each bin LOW High limit of each bin HIGH 3 5 1Swap parameter Th...

Страница 30: ...alue The compare limit value should be set in the order from small to large Figure 3 3 Tolerance mode and Sequential mode Note When setting limit values of tolerance mode the error range should be set...

Страница 31: ...ER key to input the nominal value Press 1 to input nominal value the instrument will select F H or as the default unit of the nominal value according to primary parameter 3 5 4Comparator function ON O...

Страница 32: ...w limit the DUT will be sorted into the auxiliary bin When the secondary parameter only has high limit and the auxiliary bin is set as ON if the primary parameter of DUT is within the limit range and...

Страница 33: ...the high limit of the secondary parameter 7 In Low limit of BIN 1 use numeric keys to input the low limit After inputting the data you can use p n m k M 1 to replace ENTER to input the limit value Whe...

Страница 34: ...te data time 4 1 1SKIN The function of this zone is to set the theme of display Operation steps for setting the theme of display 1 Move the cursor to SKIN the following soft keys will be displayed GRA...

Страница 35: ...s used to select short beep TWO SHT This soft key is used to select two low and short beeps 4 1 4FAIL ALARM This zone is used to control and display the beep mode as FAIL ALARM when the test result is...

Страница 36: ...t user s file MODIFY This soft key is used to modify the password The operation steps are as follows Press MODIFY to input a new password After inputting a prompt information will appear on the screen...

Страница 37: ...interface The available baud rate of this instrument is from 4 800k to 115 200k Operation steps for setting the baud rate 1 Move the cursor to BAUD RATE the following soft keys will be displayed 1200...

Страница 38: ...ess FIRMWARE UPDTE enter into the FIRMWARE UPDATE page shown as below 2 Press UPDTE the following soft keys will be displayed YES NO 3 Press YES soft key to update the firmware of the instrument 4 1 1...

Страница 39: ...ransformer scanning will be described in the part of transformer test setting Press FILE to enter into the file manage page shown as below 4 2 1Setup file for single group component EST 100 groups of...

Страница 40: ...be copied to IBM PC or compatible desk top computer laptop with USB interface to reach the infinite extension U2836 supports the USB memory devices as below Meet the USB 1 0 1 1 standard Capacity 32MB...

Страница 41: ...ps 1 Assuming that it is necessary to copy internal files 2 and 3 to files 12 and 13 respectively 2 Press FILE the file list and the following soft keys will be displayed LOAD SAVE DELETE COPY RENAME...

Страница 42: ...an 0 1 before connecting to DUT it is recommended to connect Hcur Hpot and Lpot Lcur Two terminal test In the test with high accuracy requirement using Kelvin test fixture standard accessory will gain...

Страница 43: ...o as to avoid the damage to the instrument 5 2 Eliminate the influence of stray impedance Figure 5 2 Eliminate the influence of stray capacitance When the DUT has high impedance such as small capacita...

Страница 44: ...een test leads becomes the main source of test errors except the influence of the contact resistance on test terminals If this coupling cannot be eliminated it will bring unexpected influence on test...

Страница 45: ...zone Press kHz the frequency will be changed as 5 000kHz g Move the cursor to LEVEL zone the current displayed level will be 1 000V h Press 1 5 1 5 will be displayed in the prompt information zone at...

Страница 46: ...the tested inductance is in good connection with the test fixture or not b Check the test fixture is in good connection with the test terminals of the instrument or not c Redo the open short correcti...

Страница 47: ...es Primary parameter Z Y L C R G Secondary parameter deg phase rad radian D Q RS RP G X B 6 1 3Mathematical operation Operation between the measurement value and the programmable nominal value absolut...

Страница 48: ...uency 10kHz Fast about 40 times s 27ms time Medium about 10 times s 90ms time Slow about 3 times s 300ms time The fast and middle speed will be slow down when frequency 10kHz 6 1 10 Average Be program...

Страница 49: ...01 H 99 9999kH C 0 00001pF 9 99999F Z R X 0 00001 99 9999M Y B G 0 00001 s 99 9999S D 0 00001 9 99999 Q 0 00001 99999 9 Deg 179 999 179 999 Rad 3 14159 3 14159 6 3 Measurement accuracy Test accuracy i...

Страница 50: ...Q test value 0 1 WhenDx 0 1 L C X B accuracy factor Ae should be multiplied by 2 1 x D When Qx 0 1 R G accuracy factor Ae should be multiplied by 2 1 x Q Accuracy of G can only be available in G B co...

Страница 51: ...p G combination 6 3 6 Accuracy of Rp when Dx value of tested D 0 1 The accuracy of Rp is given by the formula below Rp e x e px D D D R m Where Rpx is the value of tested Rp with the unit S Dx is the...

Страница 52: ...ure A select a smaller value on boundary line In this figure the basic accuracy value A can be obtained by the following methods 0 1 When 0 4Vrms Vs 1 2Vrms the test speed is the value A of medium slo...

Страница 53: ...V Z 100Hz fm 100kHz 200 1 10 1 3 s m V Z 70 1 10 1 9 s m V Z 100kHz f m 200kHz 200 2 10 1 3 s m V Z 70 1 10 3 9 s m V Z Fast fm 100Hz 100 1 400 1 10 5 2 3 m s m f V Z 100 1 100 1 10 2 9 m s m f V Z 1...

Страница 54: ...length 0m 1m 2m 1 5Vrms 0 2 5 10 4 1 50 fm 5 10 4 1 50 fm 1 5Vrms 0 2 5 10 3 1 16 fm 5 10 3 1 50 fm fm test frequency MHz Table E Temperature factor Ke Temperature 5 8 18 28 38 Ke 6 4 2 1 2 4 6 3 9 A...

Страница 55: ...be larger than 3 5mA AC effective value 6 5 Electromagnetic compatibility Transient sensitivity based on the requirement of GB6833 4 Conductive sensibility based on the requirement of GB6833 6 Radiate...

Страница 56: ...Standard inductor 100 H 0 02 1mH 10mH 100mH 5 Frequency counter 0 1000 MHz 6 Digital Multimeter 0 5 7 Insulation resistance meter 500V 10 levels 8 Hipot Tester 0 25kW 0 500 V 6 6 3 Function check Ens...

Страница 57: ...pectively Level 1V Range AUTO Bias 0V Speed Slow Open and short correction should be made before testing Connect standard capacitors 100pF 1000pF 10nF 0 1uF 1uF and change the frequency The error capa...

Страница 58: ...z 1kHz 10kHz 100kHz Test respectively Level 1V Range AUTO Bias 0V Speed Slow Open and short correction should be made before testing Connect standard AC resistors 10 100 1k 10k 100k and change the fre...

Отзывы: