12
Figure 5.0-2: Voltage differential test set up
If the
Model 4431-EV
is used to measure voltage differential, the CT-1 current
sensor must be disconnected and the Voltage cables connected to perform this
test. The Model 4431-EV is shipped from the factory with the capacitive sensor
programmed for the Energy test. To convert the System from energy to voltage
differential, the capacitive sensor must first be converted by plugging the red
jumpers into the center and rear pins as shown in Figure 5.0-3.
Figure: 5.0-3 Capacitive Sensor conversion from energy to voltage
The Model
4431-V
is shipped configured to the EIA 541 Voltage Differential test.
The capacitive sensor is configured for only the voltage differential
measurement.
5.2.1 R/C
Networks
The user can select either of two R/C networks using the front panel
FUNCTION selector switch. The commonly used 200pF discharged
through 1500 Ohms or the EIA 541 specified 200pF discharged through
400 kOhms.
NOTE
: The 1500 Ohm resistor will enable bags having surface
resistances between 10 and 100 kOhms (100 and 10,000 Ohms/sq) to be
differentiated. On the other hand, the 400 kOhm resistor will only be able
to differentiate bags having surface resistances between 1000 ohms and