8.0 Calibration
The HI-3627 is calibrated using a one meter diameter Helmholtz coil.
Each axis is individually calibrated. Calibration is performed at 60 Hz at
levels of 100 mG, 1 G, and 10 G. All calibrations are traceable to NIST.
The frequency response of each axis is checked in all meters.
The recommended recalibration interval is one year. ETS-Lindgren
offers complete recalibration services which comply with MIL-STD
45662A.