Model 5225 Series Shielded Enclosure

www.ets-lindgren.com

Before connecting components, read

on the back page of this document.

Safety Information

Setup Instructions

Raise the lid handle to open the door of the Model 5225.

1.

2.

(As required by the testing you want to perform) Attach cables to the SMA connectors
on the antenna patches.

4.

Place EUT inside the Model 5225 and connect cables as required.

Lower the lid handle to close the door of the Model 5225.

6.

5.

Attach a power cable to the IEC power connector.

3.

(As required by the testing you want to perform) Connect cables to the DB-9 and
fiber optic feedthroughs.

Apply power to the Model 5225. Perform testing as required.

7.

Note:

ETS-Lindgren may substitute a similar part or new part number with the same functionality

for another part/part number. Contact ETS-Lindgren for questions about part numbers and
ordering parts.

Shielding Specifications

Filter Specifications

Attenuation

100 dB, 5 MHz–6 GHz

Current

10 Amps

Voltage

100—250 VAC

Operating Frequency

0 Hz–60 Hz

Insertion Loss

100 dB, 5 MHz–18 GHz

Electrical Specifications

Weight

22.0 kg (48.6 lb)

Physical Specifications

External Dimensions

Length:

Width:

Height:

71.1 cm (28.0 in)

43.2 cm (17.0 in)

60.9 cm (24.0 in)

Internal Dimensions
(Absorber to Absorber)

Length:

Width:

Height:

60.5 cm (23.8 in)

32.3 cm (12.7 in)

50.3 cm (19.8 in)

Coupling Loss

Typical Coupling Loss Repeatability

+/- 1 dB

Typical Internal RF Absorption

2 GHz–6 GHz

Typical RF Isolation

2 GHz–6 GHz >80 dB

Better than -36 dB from

5 GHz to 6 GHz for

high frequency antennas

(~25 dB at 5.5 GHz) using

identical antennas at 10 cm

from antenna surface

30 dB to 35 dB minimum

at the 2.54 band for

low frequency antennas

Connectors

SMA feedthrough (2)

Fiber optic feedthrough (1)

Filtered DB-9 feedthrough (1)

Type N (1)

IEC power (1)