![ETROG SYSTEMS MULTIVS ES008 A Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/etrog-systems/multivs-es008-a/multivs-es008-a_instructions-for-use-manual_2434194022.webp)
CONFIDENTIAL
10/29/20 1:52:12 PM – IFU
This document may contain information that is confidential or attorney-client privileged and may constitute inside information. The contents of
this document are intended only for the recipient(s) listed above. If you are not the intended recipient, you are directed not to read, disclose,
distribute or otherwise use this document. If you have received this document in error, please notify the sender immediately and delete the
document. Delivery of this document is not intended to waive any applicable privileges
©2020 ETROG SYSTEMS LTD.
22
Recommended separation distance between portable and mobile RF communications
equipment and MULTIVS ES008A Platform
(For ME equipment ME system that are not life-supporting)
MULTIVS ES008A is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The end user of the MULTIVS ES008A
Platform can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters)
and MULTIVS ES008A as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency
of transmitter m
80 MHz to 800 MHz
𝑑
= 1.17√
𝑃
800 MHz to 2.5 GHz
𝑑
= 2.33√
𝑃
0.01
0.17
0.23
0.1
0.37
0.74
1
1.17
2.33
10
3.69
7.38
100
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
ES008A complies with the applicable requirements and relevant provisions of the Radio
Equipment Directive 2014/53/EU (RED).
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people