Doc. Ref.: ETION-MED-MDF-01-06-08 (Rev 02)
I
7
25.
CPAP
26.
PSV-S / Bilevel S
27.
PCV Bilevel T
28.
PSV Bilevel ST
29.
APCV
Refer to the
Clinical Guide
for more information
The Cube 25 NIV has been tested for the following modes:
•
CPAP – Refer to Oxygen flow measurements for CPAP mode on page 9 of this document
•
PSV/Bilevel ST (using a breath rate of 12 bpm) - Refer to Oxygen flow measurements for PSV/Bilevel ST
mode on page 9 of this document
The Cube 25 NIV must be setup in the Therapy Settings of the device for Tube Length of
1.8m / 22mm
and the
setting for the Bacterial Filter must be turned to
YES
.
The unit of measure is selectable in the Therapy Settings.
Selecting Therapy Modes and Therapy Sets
CAUTION
A physician must set the correct therapy.
CUBE 25 NIV Settings for COVID-19 Treatment
A physician is required to select the therapy
mode and program the parameters in Clinical
mode.
Maximum two
therapy sets
can be
programmed.
Once the therapy set has been set up, the
therapy set can be selected by trained staff or
the Patient.