Doc. Ref.: ETION-MED-MDF-01-06-08 (Rev 02)

 

 

Selecting Clinical/Patient Modes

 

 

 

Initial setup by Physician

 

4.

 

Decide on treatment mode 

5.

 

Select clinical mode 

6.

 

Define therapy parameters 

7.

 

Select therapy set 

8.

 

Exit clinical mode 

 

New Patient

 

9.

 

Disinfect CUBE 25 NIV device 

10.

 

Replace filter, mask etc. 

11.

 

Assemble new patient interface 

-

 

Refer to manufacturer instructions

 

12.

 

Connect to CUBE 25 NIV to test 

13.

 

Check unobstructed flow 

Treatment

 

14.

 

Place nasal prongs in patient nostrils and start 

administration of oxygen at 10L/min 

15.

 

With nasal prongs in situ, place the mask on 

patient 

16.

 

Press the Start/Stop button on CUBE 25 to 

start flow 

17.

 

Check unobstructed flow (remove plastic 

rectangular filter cover) 

18.

 

Connect CUBE 25 to mask 

19.

 

Connect second oxygen feed AFTER starting 

CUBE 25 

20.

 

Adjust flow (if required) on both oxygen feeds 

to required levels 

21.

 

Monitor patient O2 saturation, FiO2 and CO2 

22.

 

Respond to alarms as necessary 

-

 

Audio pause will silence alarm for two minutes

 

23.

 

Turn off second oxygen supply at end of therapy 

24.

 

Stop therapy on CUBE 25 after turning off second 

oxygen feed 

 

Clinical Guide 

User Guide

 

 

 

 

 

     

The device starts, in Patient mode.

 

The setting options are limited in this mode.

 

Display colours are 

Green 

or 

Purple 

in Patient 

mode.

 

 

To switch between Patient mode 

and Clinical mode, press

 

Audio Pause and OK simultaneously

 

 

    

The display colour changes to 

Blue 

,

 

The message “Clinical mode - Medical experts

 

only” appears.

 

 

 

Refer to the 

Clinical Guide 

for more information

 

Clinical Workflows

 

Содержание CHANNELS MEDICAL SOLUTONS CUBE 25 NIV

Страница 1: ...COVID 19 setup guide CUBE 25 NIV ...

Страница 2: ...lifetime of fiveyears Authorization is only provided for emergency use during the COVID 19 pandemic Etion Create intends to register a CE Mark of the CUBE 25 NIV and shall inform customers as and when this will beavailable Please contact your supplier after 31 March 2021 for expanded application clinical indication information instructions for disposal or product update options Regulatory Status a...

Страница 3: ... a Clinical and Patient mode Once the device is set up by a clinician the day to day operation can be performed by the Patient or Care Worker 4 Clinical and User Manual can be found here https www parsec co za downloads cube25 Feature CPAP PSV S PSV ST PCV APCV Range CPAP 3 to 20 hPa IPAP 5 to 25 hPa EPAP 3 to 23 hPa Target Volume 0 to 2000ml Backup rate 4 to 30 bpm Inspiration Trigger 1 to 5 step...

Страница 4: ...I 4 Doc Ref ETION MED MDF 01 06 08 Rev02 Device Interfaces For further details refer to the User Manual Device Controls Start Stop ...

Страница 5: ... used to power your Cube 25 but for backup power purposes in the event of an unexpected loss of power The UPS must be placed at least 0 5 m away from the Cube 25 and patient Be certain to plug the UPS power cord directly into a wall outlet and not into a surge protector The UPS outlets are powered whenever the UPS is switched ON During a power outage the UPS outlets will be powered for a limited t...

Страница 6: ...ructed flow remove plastic rectangular filter cover 18 Connect CUBE 25 to mask 19 Connect second oxygen feed AFTER starting CUBE 25 20 Adjust flow if required on both oxygen feeds to required levels 21 Monitor patient O2 saturation FiO2 andCO2 22 Respond to alarms as necessary Audio pause will silence alarm for two minutes 23 Turn off second oxygen supply at end of therapy 24 Stop therapy on CUBE ...

Страница 7: ... 9 of this document The Cube 25 NIV must be setup in the Therapy Settings of the device for Tube Length of 1 8m 22mm and the setting for the Bacterial Filter must be turned to YES The unit of measure is selectable in the Therapy Settings Selecting Therapy Modes and Therapy Sets CAUTION A physician must set the correcttherapy CUBE 25 NIV Settings for COVID 19Treatment A physician is required to sel...

Страница 8: ...ap 22F with cap removed to ensure leak 1963000 CMS 4 Bacterial Viral filter in exhalation 1690001 CMS 5 Oxygen adaptor for secondary oxygen feed CCSC1963 CMS 6 Anaesthetic mask Size 3 4 or 5 1514 or 1515 or 1516 CMS 7 Standard Oxygen tubing for secondary oxygen feed 1174000 CMS 8 Uninterruptable Power Supply Not shown on image UPS850LI CMS CMS Channels Medical Solutions Mask Patient Circuit Setup ...

Страница 9: ...ocument Feed 2 Secondary Oxygen feed Refer to item 7 in Mask Patient Circuit Setup on page 8 of this document NOTES This could be a bridge therapy for patients that require more than just oxygenation supplementation The oxygenation of patients should be continuously monitored Oxygen supplementation to the mask only without oxygen supply via nasal prongs seems to be not suitable for the purpose of ...

Страница 10: ...dby the lettersforming each tone pitch c a f e The C is an octave above middle C Display Priority Acoustic Cause Trigger Resolution Circuit disconnection High c a f a f Fall in pressure during exhalation Tube system probably detached 10 sec Check tube connections Mains failure Moderate 2 min Beep The device no longer has power supply Direct Check the power and UPS connections Tube blocked Moderate...

Страница 11: ...espiratory rate is below the set limit value 5 sec The patient s condition should be checked Target volume not reached Moderate c a f The target volume is not reached despite the maximum indicated inspiration pressure Immediately The patient s condition should be checked Target minute volume not reached Moderate c a f The target minute volume is not reached despite the maximum indicated inspiratio...

Страница 12: ...eaning in a well ventilated area 3 Remove coarse and fine filter 4 Use a 3 Ready to Use RBT Protectus Viridis solution Follow manufacturer instructions 5 Spray the outside of the device with RBT Protectus Viridis solution 6 Use a clean lint free cloth and wipe all obvious dirt from the device 7 Spray the device again and wait for 5 minutes disinfectant contact time 8 Connect the Cube 25 NIV to Res...

Страница 13: ...cordance with the Manufacturer s written instructions In the event of the warranty becoming void as a result of an occurrence specified in paragraph 3a to 3d above the Manufacturer will provide the Customer with a written notice of its reason for rejecting the warranty claim The Customer shall in accordance with the Supplier s written instructions deliver the defective item s at the Customer s exp...

Страница 14: ...oc Ref ETION MED MDF 01 06 08 Rev02 Authors Contributors Date Doc Ref Jaco Botha and Sanet Groenewald Andre van Tonder Andre Ten Napel Björn Kleinschmidt Jaco Basson 2020 10 14 ETION MED MDF 01 06 08 Rev 02 ...

Отзывы: