background image

For technical help, questions or comments concerning this 
product or any ETI product contact Customer Service 8:00 a.m. - 
5:00 p.m. Eastern Time.

[email protected]
networketi.com
ETI
1850 North Sheridan Street
South Bend, IN 46628

Email:

Web:

Mail:

CONTACTING CUSTOMER SERVICE

ETI’s two year limited warranty covering defects in workmanship 
and materials applies. Contact Customer Service for complete 
warranty information.

LIMITED WARRANTY

ETI makes no representations or warranties, either expressed or 
implied, with respect to the contents of this publication or the 
products that it describes, and specifically disclaims any implied 
warranties  of  merchantability  or  fitness  for  any  particular 
purpose. ETI reserves the right to revise this publication, and to 
make changes and improvements to the products described in 
this publication, without the obligation of ETI to notify any person 
or organization of such revisions, changes or improvements.

DISCLAIMER

4

Before putting the new humidity sensor on the tie wrap 

mounts, connect the two leads from the new humidity 

sensor. Reversing the process in step 6, push in on either 

orange release tab then insert either humidity sensor 

lead into that terminal. It does not matter which lead 

goes to which terminal. Release the tab to secure the 

lead. Repeat this process for the other lead. Tug gently 

on each lead to make sure it is firmly connected.

Place the new humidity sensor onto the two tie wrap 

mounts from which the original assembly was removed. 

Before securing the new humidity sensor in place, rotate 

the sensor on the two mounts so that the two air hoses 

reach  their  respective  90°  and  straight  fittings.  The 

longer hose is the Air Out hose and connects to the 

lower or straight fitting.

The shorter hose is the Air In hose and connects to the 

upper or 90° fitting. There is no right or wrong angle for 

the humidity sensor itself in terms of function. All that 

matters here is that both air hoses reach and connect 

securely to the two sensor fittings.

Once the angle of the humidity sensor is determined 

based on the position of the fittings to the hoses, secure 

the new humidity sensor in place by feeding the two tie 

wraps from the kit through the slot in each mount, 

knurled side up, wrap the tie wraps together, then cut off 

the excess from each tie wrap.

Reconnect the “Y” section of the tubing assembly which 

runs to the evaporator tray and which was disconnected 

in step 3.

As described in the Note on page 3, if any components 

were moved or removed for better access to the 

humidity sensor, replace them in their original location 

and position at this time. Restore all connections. When 

connecting the three electrical leads to the power filter 

module on AC units, make sure to connect them to the 

same terminals from which they were removed.

Reinstall the top panels using the hardware removed in 

Step 2.

Restore machine power.

7.

8.

9.

10.

11.

12.

13.

HUMIDITY SENSOR ASSEMBLY REPLACEMENT PROCEDURE  |   PART NO. 

24104 

REV 

B.0

HUMIDITY SENSOR ASSEMBLY

REPLACEMENT PROCEDURE

MODEL ADH

Отзывы: