background image

3

Shut off AC unit power by placing the power switch in the 

OFF (O) position. Unplug the power cord. Shut off DC unit 

power by shutting off the external power supply. If possible, 

move the unit to a work table.

Remove both top machine panels for easier access to the 

humidity sensor assembly. Retain mounting hardware.

Using a tubing wrench or vacuum tubing pliers, carefully 

disconnect from the “Y” hose assembly the length of hose 

running between the “Y” assembly and the evaporator tray. 

Refer back to Figure 1a for AC units and Figure 1b for DC 

units.

Using a tubing wrench or vacuum tubing pliers, carefully 

remove the two air hoses connected to the existing 

humidity sensor. Refer to Figure 2.

To replace the Humidity Sensor Assembly (23368) in either an 

ADH NETCOM Automatic Air Dehydrator with AC power or an 

ADH NETCOM with Redundant DC power, perform the steps 

below. Refer to Figure 1a or 1b, respectively, for AC and DC unit 

illustrations.

HUMIDITY SENSOR ASSEMBLY REMOVAL AND 

REPLACEMENT

1.

2.

3.

4.

Using the diagonal cutters, cut the two tie wraps securing 

the humidity sensor assembly to its tie wrap mounts then 

move the sensor out of the way to gain access to the 

electrical connector.

Using a small tool, press in on either orange release tab 

on the humidity sensor electrical connector on the PC 

board then carefully remove the lead from that terminal. 

Repeat for the other lead. Once both air hoses and both 

electrical leads have been disconnected and the two tie 

wraps cut, remove the existing humidity sensor from the 

dehydrator.

5.

6.

HUMIDITY SENSOR ASSEMBLY REPLACEMENT PROCEDURE  |   PART NO. 

24104 

REV 

B.0

HUMIDITY SENSOR ASSEMBLY

REPLACEMENT PROCEDURE

MODEL ADH

Power Filter Module and 

Switch with electrical leads 

removed (reference)

Disconnect evaporator 

tray air hose from “Y” 

assembly here. (step 3)

 Humidity Sensor Assembly

Hose to Evaporator Tray

Figure 1a. THE ADH NETCOM AUTOMATIC AIR DEHYDRATOR

HUMIDITY SENSOR ASSEMBLY IN AN AC UNIT.

Figure 1b. THE ADH NETCOM AUTOMATIC AIRDEHYDRATOR

HUMIDITY SENSOR ASSEMBLY IN A DC UNIT.

Figure 2. THE TWO HUMIDITY SENSOR AIR HOSES.

(some components removed for greater clarity)

Disconnect evaporator 

tray air hose from “Y” 

assembly here. (step 3)

Air In hose (90° fitting)

 Air Out hose

(Straight fitting)

 Humidity sensor 

electrical connector 

(on PC board)

Отзывы: