15.
Warnings
! The screw M64 [pos. 16] has no stop in the dissolving mode (left-hand rotation),
i.e. the machine support ring [pos. 01] is pressed upwards.
Remedy: Stop the machine before or turn it by hand until the gear is flat before the
new measurement process.
! When removing the transducer [pos. 04] using the eye bolt [pos. 28] it is essential to ensure
that the adapter square [pos. 05] only rests loosely and has to be removed separately.
!
If the test specimen or test device is touched during the operation of the test facility, there is
an increased risk of injury!
!
Before each commissioning, the fixed seat of the test specimen shall be checked on the test
device. Furthermore, care must be taken to ensure that the counter-holder finds a correct grip
on the start-up block. If this is not the case, a suitable start-up block must be installed.
!
Where people and objects can be harmed in the event of a break, appropriate safety measures
(e.g. covers, overload protections) must be taken by the user (observe relevant accident
prevention regulations!).
!
Note that the transducers are not designed with the usual safety factors (2...20) in favor of
high measurement sensitivity. In particular, consider the overload factors specified in the
datasheet.
!
If the test device is fitted by large temperature changes, it must not be put into operation, as
the moisture is involved in the measurement. In the case of active version, the humidity during
operation can lead to the
destruction
of the measuring amplifier.
10.
Technical data:
Please refer to the technical data in the data sheet
Test device for hydraulic tools type PH.