![ETH-messtechnik DRVL Скачать руководство пользователя страница 4](http://html1.mh-extra.com/html/eth-messtechnik/drvl/drvl_operators-manual_2432543004.webp)
messtechnik gmbh
4
Important instructions:
The model DRVL torque transducer can be deployed as a component in a test
bench, for example.
The installation position of the drive and transducer sides should be taken into
account in applications where mass is critical.
Please note that in order to achieve greater sensitivity the transducer
has not been designed with the safety factors 2 to 20 that are common in me-
chanical engineering.
Please pay special attention to the specified overload factors.
Where there is a risk of personal injury or damage to equipment due to mechani-
cal failure leading to breakage, the user of the device must put in place adequate
safety measures (e.g., covers and guards, and overload protections). All applicab-
le accident prevention regulations must be followed.
The torque transducer is not approved for use in hazardous areas.
Opening or disassembling the transducer during the warranty period
will void the warranty.