If there is an abnormality in the product, check below first.
If you use the product without familiarity with its use method, it might not work properly.
Before repair service application, check below.
7. Repair Service and Product Warranty
Problem
Checklist
Follow for solution
Low frequency
is not outputted.
Did you set the intensity to low frequency?
I did not remove the transparent film
on the pad contact face.
The pad contact face
is
dirty.
I used it immediately after washing.
The pad contact face has been damaged.
Set the intensity to low frequency
Remove the transparent film from the pad contact
face.
Wash the pad contact face with little running
water using your fingers for several seconds.
Remove water sufficiently and dry
the pad contact face naturally.
The pad is a consumable. If the pad contact face
is damaged, stop using it and purchase a separately
sold product.
Press your feet completely into the pad.
Adjust the power intensity to low/
medium frequency
Press the pad/patch/hand pad/knee pad into
the desired skin area.
Put the pad/patch/hand pad/knee pad
into the connector.
Wipe off the contact terminal of the pad/patch/
hand pad/knee pad.
Attach the pads so that they do not overlap.
Wash the pad contact face in a little running
water using your fingers for several seconds.
Adjust the intensity.
The power
intensity of
low/medium
frequency is low.
The power
intensity of the
pad’s low
frequency is low.
Did you press your feet completely into the pad?
Was the power intensity of the low/medium
frequencyset to be low?
Was the pad pressed into the
desired skin area?
Is the jack of the pad connected well?
Is the contact terminal between the pad and
cable smeared with foreign substances?
Pads are overlapped and attached.
The contact surface of the pad is dirty
The intensity is weak.
The power
intensity is strong. Was the intensity set very strong?
Adjust the intensity.
The contact
surface of the
pad is sticky.
I used it when I was sweating.
Store it in the refrigerator overnight
before use.
I frequently washed it for long time.
I put it in a place with direct sunlight,
or high temperature and humidity.
The contact
surface of the
pad is worn out.
There is sweat and water on the contact surface
of the pad.
The pad is a consumable. If the pad contact face
has been damaged, stop using it and purchase
a separately sold product.
I feel a tingling
sensation during
use. My skin
is reddened
The use time is too long.
Abide by the recommended time.
Was the area of the pad where electricity is
applied pressed completely?
Completely press the area of the pad where
electricity is applied.
Was the black area of the pad where electricity
is applied smeared with foreign substances?
Wipe off foreign substance on the black area
of the pad where electricity is applied.
The contact face of the pad is dirty or dry.
Wash the pad contact face with a little running
water using your fingers for several seconds. When
you use it, smear on the low frequency dedicated gel.
Checklist before Repair Service Application