430-11902-02, Rev M
14. TECHNICAL SPECIFICATIONS
PERFORMANCE CHARACTERISTICS
Table 14-1 lists the specifications for the System (ESU and Accessories).
Table 14-1: System Specifications
Description
Specification
Input Power
Supply Mains/Line Voltage
100
– 120 V~ (Model 4810B)
Rating
100
– 120 V~ (Model 4810BR)
220
– 240 V~ (Model 4811B)
220
– 240 V~ (Model 4811BR)
Line Current Rating
4.0 Amps, 50/60 Hz (100
– 120 V~)
2.0 Amps, 50/60 Hz (220
– 240 V~)
Fuse Rating
5.0 A, 50/60 Hz (100
– 120 V~)
2.5 A, 50/60 Hz (220
– 240 V~)
Power Cord Length
10 ft. (3 m)
Instrument Cable
Length
Connector
10 ft. (3 m)
30-pin Polarized Quick-Connect
Connector
Indifferent Electrode Cable
Current Limit
1.1 Amperes
ESU
RF Power Outputs
•
RF energy is delivered at 460 kHz,
as a sinusoid.
150 W maximum into a minimum
impedance of 37.5 Ω, with a 2 A
RMS
current limit
•
Maximum output voltage is 167
V
RMS
.
•
Maximum RF power output is based
on Probe type.
Impedance
•
Measures 25 Ω and above, if
accessing up to 150 W of RF power.
•
Measures up to 999
Ω.
Содержание COBRA 4810B
Страница 43: ...430 11902 02 Rev M Figure 14 3 Power Output versus load at full and half power Figure 14 4 Peak Output Voltage...
Страница 67: ...430 11902 02 Rev M huidplooien...
Страница 227: ...430 11902 02 Rev M condizioni operative anomale e di guasti d impostazione dell attrezzatura...
Страница 410: ...430 11902 02 Rev M 17 INFORMACI N DE CONTACTO...