44
又は
Down
8.
Joystick Calibration
1
)
Move the cursor to ‘8. Joystick Calibration’ with
or
and press
<-|
.
2
)
Make sure the cursor is positioned at ‘R_LMT’.
(The cursor is usually positioned at the place but if it is
not, move the cursor to the ‘R_LMT*’ with ‘UP’ or
‘DOWN’ switch.)
3
)
Input the correction value of R indicated on the joystick ASSY.
R correction value
4
)
Input the correction value of L, F, and B as same as 3) above.
5
)
Press “RETURN” switch 2 times. The screen will go back to the measurement screen and
the detail of calibration (values) is stored.
If you turn the power off without pressing RETURN, the detail of calibration
will not be stored. Make sure to return to the measurement screen.
Up
** MAINTENANCE MODE **
1. A r r a n g e M o d e
2. F r ee z e I m a g e
3. S h i f t S P H D a t a
4. S h i f t T a r g e t D a t a
5. Pr i n t A d j u s t D a t a
6. H a r d w a r e C hec k
7. C a li b r a t i o n
8. Joystick C a li b r a t i o n
Up
Down
<-|
Return
PLC_T
107
FBC_T
107
R_LMT
82
L_LMT
-88
F_LMT
88
B_LMT
-94
Up
Down
(+)
(-)
Return