Guidance and manufacturer's declaration – electromagnetic immunity
The equipment CSPOLA600 is intended for use in the electromagnetic environment specified below. The
customer of the user of the CSPOLA600 should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4- 6
Radiated RF
IEC 61000-4- 3
3 Vrms
150KHz to 80MHz
3V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communication
equipment should not be used close to
any part of the CSPOLA600, including
cables. Recommended separation
distance.
d=1,2*√(P)
d=1,2*√(P) 80 MHz to 800 MHz
d=2,3*√(P) 800 MHz to 2,5 GHz
Where (P) is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and (d) is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
, should be less than the
compliance level in each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note 1: at 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
CSPOLA600 is used exceeds the applicable RF compliance level above, CSPOLA600 should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating CSPOLA600.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
CSPOLA600 - Polarized visual chart system > V5 - 03-2018
18
U
SER
MANUAL
> II. S
AFETY
INFORMATION
Содержание CS POLA 600
Страница 1: ...www essilor instruments com User manual...
Страница 4: ...I INTRODUCTION...
Страница 7: ...4 CERTIFICATION CSPOLA600 Polarized visual chart system V5 03 2018 8 USER MANUAL I INTRODUCTION...
Страница 8: ...II SAFETY INFORMATION...
Страница 18: ...III FEATURES...
Страница 20: ...IV NOTES FOR USING THE INSTRUMENT...
Страница 22: ...V CONFIGURATIONS...
Страница 25: ...VI INSTALLATION...
Страница 37: ...VII TEST METHOD...
Страница 43: ...VIII REMOTE CONTROL FUNCTION...
Страница 46: ...IX HOW TO RECHARGE A REMOTE CONTROL...
Страница 48: ...X UPGRADING A CSPOLA600 SYSTEM...
Страница 50: ...XI MAINTENANCE OF UNIT...
Страница 52: ...XII TROUBLESHOOTING GUIDE...
Страница 54: ...XIII ENVIRONMENTAL CONDITIONS...
Страница 56: ...XIV SPECIFICATIONS...
Страница 58: ...XV COMPONENT LIST...
Страница 60: ...XVI SERVICE INFORMATION...
Страница 62: ...Essilor International 147 rue de Paris 94220 Charenton le Pont France www essilor com...