Installation
Installation
Installation
Installation
(1)
Do not expose the device to sunlight or bright light from other
sources.
(2)
Do not install the device in places where either dust
or rubbish may accumulate.
Also, the environments with extremes in heat and
humidity should be avoided.
In case of using the device, ensure to comply with the
environmental conditions of unpacking and usage
before starting a measurement.
・
Temperature range for use: 5
℃
to 40
℃
・
Humidity range for use: 30%HR to 95%HR
・
Temperature range for storage: -10
℃
to 55
℃
(No dew condensation)
・
Humidity range for storage: 10%HR to 95%HR (No dew condensation)
(3)
Keep away from inflammable or explosive gases as
well as storage area of the medical supplies and
chemicals.
(4)
Keep away from the sites that experience strong
vibrations or sudden shocks.
(5)
The device might be broken if it falls down. Also, it
might cause injury if dropping it. Therefore, do not
store it at an unstable place or in high, ‘out of reach’
place.
(6)
Keep this device away from water (liquid).
・
Degree of protection: IP20
Take extra caution to avoid strong light because
it may cause the failure of measurement.
NOTE
NOTE
NOTE
NOTE
Dust
Dust
Dust
Dust
Gas
Gas
Gas
Gas