5
1
3
2
Follow the same procedure (5)~(7) of the continuous keratometry and refractometry [K/R] mode.
Example of printout:
*: Confidence of result
Level
Quality
A
B
C
D
E
(Hight)
(Low)
3. Keratometry - [KER] mode
Only the radius of curvature of the cornea can be measured in [KER] mode.
Don't measure the base curve of contact lens in this mode. Measurement error will occur when it is
measured in [KER] mode.
Enter [KER] mode.
Press [Mode] button until [KER] is displayed on the right side of the display.
Follow the same procedure (2) and (3) of the continuous keratometry and refractometry [K/R] mode.
Measurement.
Press the measuring button.
Measurement will be continuously performed when the measuring button is kept pressed.
The measured results will be displayed on the monitor.
In case of the continuous measurement, previous measured result will be displayed.
When measured results are displayed, changing at PAGE 1 of [Setup] mode renders each cornea
measured result of R1/R2/AX > K1/K2/AX > AR/CY/AX displayed in turn.
U
SER
MANUAL
> M
EASUREMENT
AKR300 - Auto Kerato Refractometer > V3 - 04-2020
28
Содержание AKR 300
Страница 1: ...USER MANUAL ...
Страница 4: ...I INTRODUCTION ...
Страница 9: ...II FEATURES ...
Страница 11: ...III NOTES FOR USING THE INSTRUMENT ...
Страница 13: ...IV DESCRIPTION ...
Страница 17: ...V PRACTICING THROUGH MODEL EYE ...
Страница 20: ...VI MEASUREMENT ...
Страница 33: ...VII OTHER MODES ...
Страница 50: ...VIII SELF INSPECTION AND MAINTENANCE ...
Страница 54: ...IX CLASSIFICATIONS AND SPECIFICATIONS ...
Страница 57: ...X COMPONENTS ...
Страница 59: ...XI QR CODE ...
Страница 60: ...USER MANUAL QR CODE AKR300 Auto Kerato Refractometer V3 04 2020 60 ...
Страница 61: ...USER MANUAL QR CODE 61 AKR300 Auto Kerato Refractometer V3 04 2020 ...
Страница 62: ...USER MANUAL QR CODE AKR300 Auto Kerato Refractometer V3 04 2020 62 ...