FUNCTIONING PROBLEMS 

 
PROBLEM   

 

POSSIBLE CAUSE 

 

CORRECTIVE ACTION           

 

1.The pump doesn’t  

a. the cable is not connected 

connect the cable 

   start 
 

 

 

 

b. burnt/defective fuse 

 

change the fuse 

 

 

 

 

c.  the rotor is blocked due to 

assemble correctly 

 

 

 

 

     wrong assembly 

2. the tube is mashed  

Wrong tube dimensions or the  Use the appropriate tube or 

and the range is reduced  tube is worn   

 

 

replace it 

3. The range of the pump  a. the tube material is not   

Use the tube suitable for the 

     is incorrect 

 

    compatible with the liquid 

product 

b. dirty rolls 

Disassemble the pumping  
head and clean it  

 

 

 

 

c. wrong tube dimensions   

Use the appropriate tube 

 

 

 

 

d. The pumped product is   

Use the appropriate tube 

 

 

 

 

     very viscous 

 

4. The pumps stops  

There was an overload and Switch the pump off and    

     during work 

 

the fuse is burnt 

 

 

solve the overload cause, then 

 

 

 

 

 

 

 

 

 

change the fuse 

 

Note: security fuses are placed in the power pack at the side of the pump 

 

 

4.5

 

SWITCH OFF

 

Switch off the pump setting the switch on the position OFF. 
Disconnect the cable from the mains.  
Should the pump be kept inactive for a long period (days or weeks), remove the tube from 

the pumping head. 
If you don’t remove the compression, the tube could permanently remain deformed with a 
consequent reduction of its performances and endurance. 

Содержание IPL 10 Series

Страница 1: ...TION All the peristaltic pumps supplied by ESPANGO are mostly employed to transfer POTENTIALLY DANGEROUS chemical products STICK WITH the following instructions the assembly and the application have to be executed only by an authorized staff follow the functioning and the maintenance norms when in doubt please contact our technical dept near the pumps the operators have to wear appropriate protect...

Страница 2: ...Manuale_IPL10_ENG_R0 doc 2 INSTALLATION PROGRAMMING MAINTENANCE INSTRUCTIONS PERISTALTIC PUMPS IPL 10 RANGE IPL10 24 IPL10 30 IPL10 100 ...

Страница 3: ... 5 2 Disinfection Decontamination 6 TECHNICAL FEATURES Please carefully read the present instructions manual before starting to work with the pump and closely stick with it in order to avoid errors which could cause damages especially to the health and to the environment The data indicated in the present manual refer to the standard model Eventual variations realized on specific customer demand co...

Страница 4: ...who use the pump Please carefully read this manual before starting to work with the pump The respect of our instructions will avoid errors which could be dangerous especially concerning the health and the environment When using this pump the operator has always to refer to the present manual It has always to be available to all the people charged to use this pump We recommend to keep a copy of the...

Страница 5: ...and non flammable surface and be sure that it is firmly supported The pump has to be protected from direct sunlight UV radiation 2 2 ELECTRICAL CONNECTION The pump has a flexible cable Please use the plug to connect and disconnect the pump from the mains Before switching the pump on verify that the cable and the plug are not damaged in case they are do not connect the pump to the mains The voltage...

Страница 6: ...r 3 roller rotor 4 tube 4 tube 5 plastic fittings 5 plastic hooks IPL 10 100 head IPS100 Components identification 1 moving part 2 roller rotor 3 fixed part On request and for special applications different pumping heads models can be used In this case make reference to the technical indications given by our technical dept 1 2 3 4 5 8 6 9 7 3 2 CONTROL PANEL 1 Operation light optional foot switch ...

Страница 7: ...S100 1 Open the upper mobile part of the head 1 gently pulling it upward 2 Withdraw the old tube 3 Lay the new tube on the roller rotor 2 inside the moving head keeping it in slight tension at its ends 4 Still keeping the tube in slight tension close the upper part of the moving head until you hear the coupling click Now the pump is ready for use IMPORTANT NOTE we guarantee its correct functioning...

Страница 8: ...lues How to activate the pump select on the front from the key no 1 the position I or II on the position I the timer will be activated by the pulse of the pedal press the pedal ONCE or press the key 3 START on the control panel The pump will work for the selected uptime and will stop at its expiry releasing a beep The pump is now waiting for a new input to run again In order to PAUSE the pump DURI...

Страница 9: ...ty rolls Disassemble the pumping head and clean it c wrong tube dimensions Use the appropriate tube d The pumped product is Use the appropriate tube very viscous 4 The pumps stops There was an overload andSwitch the pump off and during work the fuse is burnt solve the overload cause then change the fuse Note security fuses are placed in the power pack at the side of the pump 4 5 SWITCH OFF Switch ...

Страница 10: ... to proceed to an accurate cleaning of the pumping head after a long running period Residual particles coming from the normal wear of the tube could adhere to the rollers compromising their precise function Product residuals have to be regularly removed from the suction and delivery fittings so that a correct flow of the product is assured avoiding bottlenecks The motor module and the head bearing...

Страница 11: ...th mm 146 x 120 x 240 IPS30 width x height x length mm 166 x 120 x 240 IPS100 width x height x length mm 175 x 120 x 240 Weight kg 1 5 230 Volt version Power V Hz 230 Volt 50 60 Hz W head IPS24 10 Watt head IPS30 30 Watt head IPS100 50 Watt Fuse A V head IPS24 1 x Fuse 1 5 A 250 V delayed IPS30 1 x Fuse 1 5 A 250 V delayed head IPS100 1 x Fuse 2 0 A 250 V delayed 24 Volt AC DC version Power Fuse V...

Отзывы: