EsiWelma Srl 

URS20SE - EW0826A1_en.docx 

Oxygen detectors – URS20SE 

16/06/2017 

Gas detection systems for industrial environments 

1/8

 

 

 

 

 

 

 

 

Oxygen detectors 

ATEX certified (II 2G Ex d IIC T6) 

 

URS20SE 

 

 

Sssssss 

 

 

 

 

 

 

 

11

…28Vdc power supply. 

Electrochemical cell sensor designed for the detection of oxygen (O

2

). 

Up to three alarm thresholds plus sensing element fail. 

LED on sensing element body to indicate operating status. 

Automatic countdown of sensor life. 

 

Use 

The  URS20SE  detectors  are  used  to  detect  the  presence  of  oxygen  (O

2

),  in  areas 

classified as Zone 1. 

An

  Oxygen  excess

  forms  if  it  leaks  in  hospitals,  laboratories,  welding  centres  and, 

more in general, where oxygen is stored. 

Oxygen deficiency

 is an indirect measurement of the presence of other explosive or 

asphyxiant gases that deplete the oxygen in the air. 

URS20SE sensors can be used in stand-

alone mode with 4…20mA output or with an 

optional voltage-free contact relay card having the following four digital outputs: 

Pre-alarm, 1

st

 alarm threshold, 2

nd

 alarm threshold, sensor fail. 

Operation 

If  there  is  an  oxygen  leak  or  an  absence  of  oxygen,  the  detector  compares  the 
measured  concentration  value  with  the  threshold  limit  setpoints  and  energises  the 
associated  relays.  Information  on  the  measured  concentration  value  is  always  at  the 
4…20mA output. 

Ordering 

To order, simply state the part number: 

URS20SE

For special versions, on request, please contact Customer Service. 

 

Содержание Sensigas URS20SE

Страница 1: ...entres and more in general where oxygen is stored Oxygen deficiency is an indirect measurement of the presence of other explosive or asphyxiant gases that deplete the oxygen in the air URS20SE sensors...

Страница 2: ...default Consumption mode applications at 1 or 2 thresholds 4 20mA Output reference selection 4mA 0 O2 20mA 30 O2 0mA no alarm 10mA 1st threshold alarm 20mA 2nd threshold alarm by jumper selectable pol...

Страница 3: ...n this case the detectors must be located at the breathing height of the occupants in the premises Take into consideration the following specific installation guidelines as well as the above instructi...

Страница 4: ...iameter of the cable gland Default settings of the sensor are shown in the Technical Specifications chapter In order to change the default settings switch off the power supply input the new settings a...

Страница 5: ...ions below Step 1 Insert the connection slot provided with the relay card into the control card making sure the flexible tab is towards the main terminal board Find CN3 connector CN3 connector Connect...

Страница 6: ...factory calibrated so they normally do not require any other calibration once installed Still after installation an operational check of the sensors is recommended The detector will enter a 2 minute w...

Страница 7: ...rned to the supplier installer who in turn will send it back to the manufacturer Sensors may need to be re calibrated using the TUL40 calibration kit and the TUS40 20 handheld terminal which must be c...

Страница 8: ...industry 2 Equipment category 2 for use in Zone 1 G Equipment intended for use in explosive gas atmosphere caused by mixture of air and gas vapours flammable mist Ex d IIC T6 Type of protection compl...

Отзывы: