ESD SYSTEMS 41290 Скачать руководство пользователя страница 2

PS-2111

Page 2 of 7

ESD protective products
should be tested:

A. Prior to installation to qualify for

listing in user's ESD control plan
approved ESD protective materials

B. During initial installation
C. For periodic checks of installed

products as part of ESD control plan
compliance verification plan

“A Compliance Verification Plan shall
be established to ensure the
Organization’s fulfillment of the
technical requirements of the ESD
Control Program Plan. Process
monitoring (measurements) shall be
conducted in accordance with a
Compliance Verification Plan that
identifies the technical requirements to
be verified, the measurement limits and
the frequency at which those
verifications shall occur. The
Compliance Verification Plan shall
document the test methods and
equipment used for process monitoring
and measurements. If the test methods
used by the Organization differ from
any of the standards referenced in this
document, then there must be a tailoring
statement that is documented as part of
the ESD Control Program Plan.
Compliance verification records shall be
established and maintained to provide
evidence of conformity to the technical
requirements.

The test equipment selected shall be
capable of making the measurements
defined in the Compliance Verification
Plan.” (ANSI/ESD S20.20-2007 section
7.3)

Electrification Period

The Surface Resistance Test Kit
provides the proper electrification
period of 15 seconds per ESD S4.1,
after numerous readings and
calculations are executed, then displays
surface resistance mantissa
measurement (Note: most analog type
meters display measurements
instantaneously).

When the Test Button is depressed, the
liquid crystal display (LCD) will
indicate:

• Temperature in degrees Fahrenheit

(tolerance ±5°F, typical)

• Temperature in degrees Celsius

(tolerance ±3°C, typical)

• Humidity as percentage (from 5% -

95% tolerance, ±10 typical) 

• Surface resistance mantissa (with

exponent displayed via LED,
measurement in ohms)

Measurement being displayed is
identified by the illuminated function
LED. The surface resistance exponent
(or power of number) is immediately
illuminated and remains illuminated
measuring the range the surface
resistance in ohms.

Reference Literature

In addition to those noted above:
ANSI/ESD S20.20 - Development of
ESD Control Program
ESD ADV1.0 - Glossary of Terms
ANSI/ESD S6.1 Grounding

These documents can be obtained
directly from the ESD Association, 7902
Turin Rd., Suite 4, Rome, NY 13440-
2069, (315) 339-6937 or www.esda.org.
Other standards are available from the
agencies who produce them. MIL-
HDBK 263A, EIA-IS-5-A,
ASTM-F-150, EN 100015, and EIA-
625.  If you need help in obtaining these
documents, please contact our customer
service department.

Features

A. LED Displays:

Surface resistance

exponent is displayed via light emitting
diodes illuminating 1/8" high number.
There are 12 LEDs displaying surface
resistance exponent measurement. They
are color coded for quick checks:

Exponent

Color

<3, 3

Red

4, 5

Green

6, 7, 8

Blue

9, 10, 11

Green

12, >12

Yellow

Five Function LED’s identify the
measurement being displayed (see B
below).
• When battery voltage drops to

approximately 2 volts, one of the
Function LED's to the right of the red
pushbutton switch will begin to flash,

indicating the need to replace
batteries.

A

B

C

D

E

F

G

© 2007 DESCO INDUSTRIES INC.

Employee Owned

B. Liquid Crystal Display (LCD):

Mantissa is displayed via 9/16" high 

liquid crystal  display and provides easy
to read resistance (or resistivity)
measurements.

When the Test Button is depressed, the
LCD will indicate:
• Temperature in degrees Fahrenheit
• Temperature in degrees Celsius 
• Humidity as percentage 
• Surface resistance mantissa (with

exponent displayed via LED,
measurement in ohms)

Surface resistance ohm values are
expressed with a mantissa and exponent
or power of the number. For example, if
"8" is illuminated by its LED and the
LCD displays a mantissa of "7.14", the
measurement is 7.14 x 10E8 ohms or
714,000,000 ohms

If surface is over 10E13 ohms, the
following will appear:
• "1____" on the LCD display indicate

Overrange or that the reading exceeds
the display capabilities

Figure 2. Features of the Surface
Resistance Meter

Содержание 41290

Страница 1: ...loor Footwear ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person Workstations ESD ADV53 1 ESD Protective Workstations Versatility Measures RTT RTG or Resistivity requires optional Resistivity Attachment item 41292 Quick Checks Surface resistance exponent number illuminates immediately i e 8 10E8 ohms or 100 000 000 ohms Periodic Checks of Installed Product...

Страница 2: ...ing displayed is identified by the illuminated function LED The surface resistance exponent or power of number is immediately illuminated and remains illuminated measuring the range the surface resistance in ohms Reference Literature In addition to those noted above ANSI ESD S20 20 Development of ESD Control Program ESD ADV1 0 Glossary of Terms ANSI ESD S6 1 Grounding These documents can be obtain...

Страница 3: ...urfaces use ESD Systems com Reztore Antistatic Surface and Mat Cleaner item 10630 or other ESD cleaner not containing silicone Be sure the surface is dry before testing Periodic maintenance The area surrounding the cable jacks at the top end of the meter should be wiped with a clean cloth moistened with alcohol to remove skin oils that will accumulate and affect the accuracy at high resistances Th...

Страница 4: ...ulative silicone i e Reztore Antistatic Surface and Mat Cleaner Reporting and Using Test Results Different standards have different requirements use requirements as specified in user s ESD control plan Examples are Per ESD 4 1 worksurfaces Report RTG maximum and minimum values measured for resistance to ground RTT maximum and minimum values measured for point to point resistance in ohms Per ESD 7 ...

Страница 5: ...ing Diodes LED s illuminate exponent numbers and function being performed Accuracy Resistance measurements within 10 E12 and greater 20 complies with ESD S4 1 Open circuit voltages of 10 volts 5 and 100 volts 5 exceeds requirements of ESD S4 1 Meter Weight 15 5 oz Dimensions 8 L x 4 3 W x 1 6 H Test Switch Press the red test button for 15 20 second electrification period Meter will immediately dis...

Страница 6: ...o AUTO switch up Testing each decade starting from 1 0 x 10E12 and down never start from 1 0 x 10E3 Temperature Fahrenheit 75 F 6 6 Celsius 23 8 C 10 2 Relativity Humidity 10 Digits A 1 0 x 10E12 20 LED 12 Yellow Mantissa 1 20 0 LED 12 Yellow Mantissa 1 00 20 LED 11 Yellow Mantissa 8 00 B 1 0 x 10E11 20 LED 11 Yellow Mantissa 1 20 0 LED 11 Yellow Mantissa 1 00 20 LED 10 Green Mantissa 8 00 C 1 0 x...

Страница 7: ...390 for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The expr...