Escort ChartReader EA-CR Скачать руководство пользователя страница 7

USER GUIDE

 

 

 

©

 

Escort Data Logging Systems LTD         DATED: 03/07/07                PAGE 7

 

 

 

L

 

Multiple Downloads - Example

 

 

The following is an example of how to use the multiple logger download feature of the 
ChartReaders with 1MB FLASH memory. The example is based on a warehouse with several 
cool rooms that are furnished with one logger each to record the temperature. On a regular 
basis the graph or out of specification readings have to be printed.  

 
Note:

 Multiple downloads is enabled by default. To deactivate this feature please use the 

ESCORT user software and select ‘N’ to the ‘Enable Multiple Downloads’ option in the 
configuration settings (refer to the help menu). 
 

Example 

1.  Perform a battery check by pressing the 

  +  

  keys. The battery condition should 

be OK or better. If not then the battery should be charged before proceeding. 

2.  Print a data block summary by pressing the 

  +  

  keys. 

3.  If there are data blocks stored in memory then delete them by pressing the 

  +  

  keys. If there are more than one data block to be deleted, the 

  key can 

be held down and the 

 key repeatedly pressed on and off until all the data blocks are 

deleted. After each data block is deleted the ChartReader will produce a single beep. 
When there are no more data blocks left to delete the ChartReader will produce a series 
of beeps. 

4.  Now the ChartReader is ready to be used in the warehouse. For each logger :- 

4.1. 

Make sure the ChartReader is turned ON (the active indicator light will be ON if 
the ChartReader is turned ON ). 

4.2. 

Connect the logger to the ChartReader interface and press the 

 key. The 

ChartReader will download the logger’s readings into a data block. 

4.3. 

At this time the logger’s readings could be printed out via pressing the chart 

 

key or out of specifications 

 key. 

4.4. 

If required the logger could be re-armed (programmed anew) by pressing the 

  

key. This would stop and set up the logger for a magnetic start with the same 
logging parameters as before. 

4.5. 

Remove the downloaded logger from the ChartReader interface, and if it was re-
armed place it where required and swipe the start magnet across the logger. 

4.6. 

At any stage a data block summary 

(

 

    +   

 

)

 or data block list 

(

 

  +  

 

)

 can be printed out to verify which loggers have been downloaded. 

5.  At a convenient time/place print a data block list with the 

  +  

  keys. This will 

show you the order of how the loggers where stored in your ChartReader. 

6.  The last data block is always the one that is active for printing. So if it needs to be printed 

in chart form or out of specifications only then press the 

 or 

 keys respectively. 

7.  Once the data block has been printed, delete it by pressing the 

  +  

  keys once. 

CAUTION: Once a data block has been deleted it can not be retrieved. Pressing the 

  +  

  keys more than once will delete data blocks that may not have been 

printed. 

8.  Now that the data block that was printed in step 6 has been deleted, the previous data 

block has now become the last data block and can now be printed in the same way. 

9.  Continue printing and deleting the data blocks until all data blocks have been deleted. 
 

Содержание ChartReader EA-CR

Страница 1: ...er EJ CR II EJ CR MEM MN CR MEM EA CR EI CR MEM User Manual Thank you for purchasing an ESCORT dls product Please read this user manual before using this product www escortdls com ESCORT Data Logging...

Страница 2: ...y Check Function 3 Latest CPU Technology 4 Flash Memory Version 4 Charts 8 Out of Specification Readings 9 Powering Charging the ChartReader 10 Power Failure 10 Overview of Power Indicator Modes 10 Te...

Страница 3: ...ser software and select N to the Enable Multiple Downloads option in the configuration settings refer to the ESCORT software user manual and help menu Logger Variants All variants in the Junior produc...

Страница 4: ...ery fails before the user prints the logger information or downloads the information to a computer system This should not be a problem since the user can check the battery status at any time However t...

Страница 5: ...beeps are sounded Chart Printing To print a chart or graph of logged data press the chart key This will print from the latest data block stored in memory the last logger read The chart is created acc...

Страница 6: ...g table shows available conditions Condition Action Excellent Do not charge the battery Good Ready for general use OK Ready for light to medium use Fair Should be charged Poor Definitely charge before...

Страница 7: ...indicator light will be ON if the ChartReader is turned ON 4 2 Connect the logger to the ChartReader interface and press the key The ChartReader will download the logger s readings into a data block...

Страница 8: ...in relation to upper lower specification and increment values or just full scale and can be set using the ESCORT user software Times are Sender local Frequency of Date Time stamp Timemarks are select...

Страница 9: ...If the logger has two sensors such as a EJ IE or EI HS loggers then the other sensor s specifications and log summary will also appear if the sensor is enabled for logging For EJ IE EJ HS loggers the...

Страница 10: ...to fully charge a battery Note It is highly advised that the battery is fully exhausted before recharge to maximise life and capacity of the battery Note that because the ChartReader uses an intellige...

Страница 11: ...nd EJ CR MEM iLog ChartReaders currently support the following ESCORT iLog products EI IN D 32 L EI 1E D 32 L EI 1E DC 32 L EI 1E DN 32 L EI 1E DZ 32 L EI 2E D 32 L EI HS D 32 L Supported Junior Tempe...

Страница 12: ...l and help menu Below please find an overview of the options available SITE OR RECEIVER DESCRIPTION Up to 4 lines of 24 characters of text are available for general purpose This text appears at beginn...

Страница 13: ...rough the mechanism by hand in normal direction PAPER LOADING PATH Feed through Approx 6mm Ribbon Cassette p Remove old roll core from the holding fingers and insert the new roll with the paper direct...

Страница 14: ...ader o Press down on the right end of the ribbon cassette to pop the left end up The cassette can now be lifted out p Pass the paper through the new ribbon cassette and load the cassette in the mechan...

Страница 15: ...s Printer type Epson M190 impact dot matrix printer Printout graph size Paper size 57 mm Y range 46 mm 1 7 C mm full scale 40 to 40 C X range 37mm 100 logs Printout Speed typical 3 5 lines per second...

Страница 16: ...ronmental conditions outside those deemed acceptable by EDLS In case of a warranty claim ESCORT Data Logging Systems Ltd will repair the goods or supply an equivalent replacement In some circumstances...