ets-lindgren.com
Series H-491198-01 Battery Charger
77
Following the maximum time of 180 minutes, a pulse-trickle charge mode is used
to compensate for the self-discharge of the batteries while idle in the charger.
The fast charge current is low enough that there is not always enough heating of
the battery cells to cause the voltage to drop. With no voltage drop there is no
peak voltage detection, and the fast charge will terminate after the maximum time
of 180 minutes is reached.
S
YNCHRONIZING THE
B
ATTERY
C
HARGE
I
NDICATOR
The probe has internal circuitry to track the charge condition of the batteries.
NiMH batteries have a sharp discharge-curve knee, which means there is little
indication that the batteries are nearly discharged prior to a low battery condition.
The internal circuitry tracks the charging and discharging currents of the batteries
and displays the calculated battery condition.
It is possible that the battery charge indication may lose synchronization with the
actual battery condition. A loss of synchronization may be indicated by a low
battery charge alert after the batteries were recently fully charged. Loss of
synchronization may also be indicated by short battery life and premature low
battery alert.
To synchronize the battery charge indicator:
1.
Make sure the probe is off or the batteries will not charge.
2.
Connect the battery charger to the electrical mains, and then plug the
charger output into the battery charger connector on the probe.
3.
Press the
DISCHARGE
button on the battery charger. The batteries
will fully discharge, and then fully recharge. The total time required to
discharge and recharge is approximately seven hours.
4.
For best results, repeat the process once. However, it may be
necessary to repeat the process three times for proper
synchronization. The synchronization process may take up to 12
hours.
When completely charged, the battery indicator for the probe should indicate a
charge level of more than 90%.
Содержание ETS-Lindgren HI-6006
Страница 1: ...HI 6122 HI 6105 HI 6153 HI 6023 HI 6006 HI 6053 EMC Field Probes User Manual ...
Страница 10: ...x ets lindgren com This page intentionally left blank ...
Страница 20: ...20 Typical Configurations ets lindgren com This page intentionally left blank ...
Страница 34: ...34 Maintenance ets lindgren com This page intentionally left blank ...
Страница 44: ...44 Specifications ets lindgren com This page intentionally left blank ...
Страница 61: ...ets lindgren com Typical Data Laser Powered Probes 61 Typical Isotropic Response in dB at 10 GHz ...
Страница 62: ...62 Typical Data Laser Powered Probes ets lindgren com Typical Isotropic Response in dB at 18 GHz ...
Страница 69: ...ets lindgren com Typical Data Battery Operated Probes 69 Typical Isotropic Response in dB at 10 GHz ...
Страница 70: ...70 Typical Data Battery Operated Probes ets lindgren com Typical Isotropic Response in dB at 18 GHz ...
Страница 72: ...72 Warranty ets lindgren com This page intentionally left blank ...
Страница 138: ...138 ETSProbe DLL User Guide ets lindgren com This page intentionally left blank ...