Specifications
Power
Main:
IEC filtered AC power input module ; 110-240 VAC, 500 mA
max, 50-100 Hz
Alternate:
Automobile cigarette lighter to 2 mm power plug adapter
cord, 12.5 Vdc, 100 mA
Fuses:
250 Volt, 1.0 Amp, Type T (5 mm x 20 mm)
Output
Open Circuit Voltage:
15 Vdc
Fast Charge Pending Current:
60 mA
Fast Charge Current:
1400 mA
Pulsed Trickle Charge Current:
50 mA
Output Voltage (During Fast
Charge):
3–6 Vdc
Battery
NiCd Battery:
3.6 Volt 3 Cell NiCd Battery, 1400 mAH (rapid charge cells, 1.2
Volts/cell)
ETS-Lindgren Part #491038
Environmental
Operating Temperature:
10°C–40°C (50°F–104°F)
Humidity:
5% to 95% relative humidity, non-condensing
Maintenance Recommendations
•
Operate the battery charger with care.
•
There are no user serviceable parts inside the battery charger. Opening the
battery charger housing may void your warranty.
Replacing the Fuse
If the battery charger fails to operate, check for a blown fuse inside the power
entry module. A blown fuse must be replaced with the same value and type of
fuse, or an unsafe condition may result. Use only 250 Volt, 1.0 Amp, Type T (5 mm
x 20 mm) fuses.
To replace a fuse:
1. Two fuses are located in the fuse drawer in the power input module. Use a
screwdriver to open the drawer.
2. The fuse towards the outside of the drawer is the spare. Remove the spare
fuse from the module.