ESCO Technologies ETS-Lindgren HI-3702 Скачать руководство пользователя страница 15

15

14

Never attempt to 
recharge a non-
rechargeable battery.

Note

If the battery has been 
through its normal life 
span of approximately 
1000 charge/discharge 
cycles, return the HI-
3702 to ETS-Lindgren 
for battery replacement 
and recalibration. For 
more information, see 
Battery Maintenance 
on page 9.

Displayed readings that are less than 10% of the full scale reading for the selected 
range are not valid. When 

Range 1 

or 

Autorange

 is selected and there is no 

current passing through the current meter, a noise floor reading will display. This 
is a normal operating condition.

Computer Controlled Data Logging

Software is available for the personal computer that simulates the HI-4416. For 
information on using the HI-4416 for data logging, see the HI-4416 manual.

The HI-3702 is 
connected for data 
logging as shown.

To connect the HI-3702 to a PC serial port, use the optional model HI-4413P and 
RS232 cable assembly (PN H-2239615). For connection to a PC USB port, use the 
optional model HI-4413USB (cable included).

Battery Charging

1.  Turn off the HI-3702.
2.  Connect the battery charger/wall transformer to an AC outlet.
3.  Plug the charger jack completely into the CHARGE connector on the clamp 

on current sensor.

       Allow approximately one hour for a full charge cycle. When charging is   
       complete, the battery charger automatically goes into a trickle charge. It       
       will remain in trickle charge mode until disconnected from the HI-3702.
4.  When the charge cycle is complete, remove the charger jack from the 

CHARGE connector on the clamp-on current sensor.

5.  Unplug the battery charger from the AC outlet.
6.  Use the HI-4416 to verify that the battery is charged to approximately 3.8 VDC.

See the appendices for your model for complete information on the battery 
charger.

Содержание ETS-Lindgren HI-3702

Страница 1: ...Model HI 3702 Clamp On Induced Current Meter User Manual PN H 600070 August 2022 Rev J ets lindgren com...

Страница 2: ...of this document may be copied by any means without written permission from ETS Lindgren Inc Trademarks used in this document The ETS Lindgren logo is a registered trademark of ETS Lindgren Inc Teflo...

Страница 3: ...cs 10 Replacement and Optional Parts 10 Service Procedures 10 RANGES SPECIFICATIONS 11 Ranges 11 Specifications 11 HI 3702 Typical Frequency Response 12 OPERATION 13 Computer Controlled Data Logging 1...

Страница 4: ...R 21 Introduction 21 Charging the Battery 21 Charging Indicators 21 Specifications 22 Maintenance Recommendations 22 APPENDIX C EC DECLARATION OF CONFORMITY 23 Hi 3702 Clamp On Induced Current Meter 2...

Страница 5: ...y damage Included text gives proper procedures WARNING Denotes a hazard Failure to follow instructions could result in SEVERE personal injury and or property damage Included text gives proper procedur...

Страница 6: ...ble power cable or autotransformer without a protective ground conductor Caution Uninsulated voltage within the unit may have sufficient magnitude to cause electric shock Therefore it is dangerous to...

Страница 7: ...ures at 3 kHz to 200 mA for controlled exposures between 100 kHz and 100 MHz The accuracy of the HI 3702 provides a means for determining compliance The unit is completely portable and self contained...

Страница 8: ...roduct Information Bulletin included with your shipment for the following Safety regulatory and other product marking information Steps to receive your shipment Steps to return a component for service...

Страница 9: ...ould the unit be immersed in water or other liquid the HI 3702 is not waterproof Such actions void the warranty Battery Maintenance Maintaining the HI 3702 requires that the nickel cadmium NiCd batter...

Страница 10: ...ntegrity Before connecting fiber optic cables clean the connector tips and also the in line connectors Before attaching in line connectors clean them with moisture free compressed air Replacement and...

Страница 11: ...200 micron Graded Index Multimode Max Length 1 Km Readout HI 4416 Digital Readout Control Unit Physical Specifications HI 3702 Inner Diameter 105 4 mm 4 15 in HI 3702 Outer Diameter 171 4 mm 6 75 in H...

Страница 12: ...12 HI 3702 Typical Frequency Response...

Страница 13: ...cables Connect the fiber optic cable to the HI 3702 and to the HI 4416 Digital Readout Control Unit Follow the color coding on the fiber optic cable the HI 3702 and the HI 4416 3 Power on the HI 3702...

Страница 14: ...A 6 Select averaging Use the HI 4416 to select Instantaneous Readings or One Second Averaging The HI 3702 collects a measurement approximately seven times per second When the X axis is selected each m...

Страница 15: ...onnected for data logging as shown To connect the HI 3702 to a PC serial port use the optional model HI 4413P and RS232 cable assembly PN H 2239615 For connection to a PC USB port use the optional mod...

Страница 16: ...rriage return Error String E01 Communication error for example overflow E02 Buffer full error too many characters contained between the start character and carriage return sequence E03 Received comman...

Страница 17: ...es not perturb the measured RF field or the current distributions The sensor is a ferrite split core current transformer The signal generated by the sensor is amplified by 10 dB to 50 dB depending on...

Страница 18: ...splays LEDs that provide the operating status The battery charger connects to the device being charged through a short cord terminated with a power jack An integrated circuit within the battery charge...

Страница 19: ...38 Environmental Operating Temperature 10 C 40 C 50 F 104 F Humidity 5 to 95 relative humidity non condensing Maintenance Recommendations Operate the battery charger with care There are no user servic...

Страница 20: ...20 3 Replace the blown fuse with the spare fuse 4 Slide the fuse drawer back into the module Make sure that the drawer snaps securely into its locked position...

Страница 21: ...connects to the device being charged through a short cord terminated with a power plug An integrated circuit within the battery charger monitors the battery voltage and controls the charging function...

Страница 22: ...c Battery NiCd Battery 3 6 Volt 3 Cell NiCd Battery 1400 mAH rapid charge cells 1 2 Volts cell ETS Lindgren Part 491038 Environmental Operating Temperature 10 C 40 C 50 F 104 F Humidity 5 to 95 relati...

Страница 23: ...23 APPENDIX C EC DECLARATION OF CONFORMITY Hi 3702 Clamp On Induced Current Meter...

Страница 24: ...24 Series H 491198 36 Battery Charger...

Страница 25: ...25...

Отзывы: