MIRI
®
TL family`s multiroom IVF incubators User Manual Rev. 1.0
46
These items are placed in the MIRI
®
TL12 multiroom IVF incubator menu, and they are
named: T1, T2, T3, T4, T5, T6, T7, T8, T9, T10, T11, T12 T13, T14, T15, T16, T17, T18,
T19, T20, T21, T22, T23 and T24.
An overview of the areas associated with the sensor names is shown in the tables below.
Table 15.1
Areas associated with sensors in the MIRI
®
TL6 multiroom IVF incubator
Area
Bottom
Lid
Compartment 1
T1
T7
Compartment 2
T2
T8
Compartment 3
T3
T9
Compartment 4
T4
T10
Compartment 5
T5
T11
Compartment 6
T6
T12
Table 15.1
Areas associated with sensors in the MIRI
®
TL12 multiroom IVF incubator
Area
Bottom
Lid
Compartment 1
T1
T7
Compartment 2
T2
T8
Compartment 3
T3
T9
Compartment 4
T4
T10
Compartment 5
T5
T11
Compartment 6
T6
T12
Compartment 7
T13
T19
Compartment 8
T14
T20
Compartment 9
T15
T21
Compartment 10
T16
T22
Compartment 11
T17
T23
Compartment 12
T18
T24
To calibrate temperature in a particular area, please find the corresponding sensor name
and adjust it according to the measurement taken using a high-precision thermometer.
Temperature calibration is done by adjusting the Tx (where x is the sensor
number) according to the measurement done on the spot relevant to the dish
placement.
After temperature adjustment, give it at least 15 minutes for the temperature
to stabilize, use the thermometer to verify the correct temperature on each area.
Be careful when changing the calibration settings
–
make sure that only the altered value
corresponds to where the measurement is done. Give the system some time to adjust.