Figure 7.3
No contact between the floating plate and the table
It is very important that there would be no contact between the floating plate and the
table. Then place the float element 2 on the top of float element 1.
Figure 7.4
Rubber stops facing up
The rubber stops should be facing up. Leave equal distance from the hole on all sides
of the table plate.
Then place the top of the floating element (stainless steel) on the rubber parts again,
leaving equal distance from the hole on all sides of the table plate.
Figure 7.5
Float element is placed too high
The float element should be level with the table top. In the picture above it is too high.
The adjustment can only be done after the microscope has been placed on the float, as
the weight will push it down.