Where:
OUT
indicates that the instrument is performing a generation.
AUTO indicates the autoranging function.
XXXXX it shows the programmed value.
mV
is the engineering unit.
By the numerical keyboard, set the value to be generate.
Push the FUNC push-button, the instrument start immediately to
generate.
3.11.2 - RNG(L) mV generation with linear read-out scaling.
The
RNG(L)
selection allows to associate to a read-out range a
desired linear output range.
Example 1
: with a read-out going from -100.0 to 1000.0 (mB) it is
desired to generate a linear output from 0 V to 10 V.
Pushing
s
or
t
push-button, select the RNG (L) function.
Push the FUNC push-button, the display will show the following:
By the numerical keyboard, set the desired initial scale value (For
Example 1, this value is equal to 0 mV).
NOTES
:
1) The value assigned to "H.RANGE mV" must be higher than the
value assigned to "L.RANGE mV" parameter.
2) The value assigned to "H.RANGE EU" parameter must be
different from the value assigned to "L.RANGE EU".
When conditions 1 or 2 are not satisfied, the instrument will reject
the new value showing the "0" and it will maintain the cursor on the
rejected parameter.
3) The value assigned to "H.RANGE EU" parameter can be lower
than the value assigned to "L.RANGE EU" (reverse read-out).
Push the FUNC push-button; the cursor will go automatically on the
lower line.
By the numerical keyboard, set the desired read-out assigned to
the initial scale value (For Example 1, it is equal to -100.0 mB).
Depress the FUNC push-button.
NOTES
: For the read-out in eng. units, the number of decimal
figures is free with respect of the read-out limits (20000 counts).
The display will show the following:
By the numerical keyboard, set the desired final scale value (For
Example 1, this value is equal to 10000 mV).
Push the FUNC push-button.
By the numerical keyboard, set the desired read-out assigned to the
final scale value (For Example 1 this value is equal to 1000.0 mB).
L.RNG XXXXXX mV
L.RNG XXXXXX EU
H.RNG XXXXXX mV
H.RNG XXXXXX EU