39
CONNECTION FOR TC SIMULATION USING EXTERNAL RJ
When it is desired to simulate a TC and an external stable refer-
ence junction box is used, connect the copper cable of the stable
reference junction box to the instrument and set the external
reference junction compensation value as show below.
HOW TO SIMULATE A TC SENSOR
Depress the OUT push-button, the display will show the following:
By
s
or
t
push-buttons select the "TC" sensor.
Push the FUNC push-button; the display will show the following:
By depressing the
s
or
t
push-buttons select the desired TC type
(e.g. J).
NOTE
: when ITS-90 is selected, the TC type U and L are not
available so that the instrument will show these two TC types with a
small letter and it does not allow to select them by jumping the two
symbols during the cursor scrolling.
Depress the FUNC push-button again the display will show the
following:
By pushing the
s
or
t
push-buttons select the desired engineering
unit (e.g. °C).
+
-
copper cable
external reference
junction box
TC extension cable
TC
MEASURING
INSTRUMENT
OUT
TC RTD mA mV
K J L E T U W W3
S R B N P Nm W5
ENGINEERING UNIT
°C °F
M20-e-3.p65
7/9/01, 2:40 PM
39