Captair Bio Smart

User manual

Mobile ductless filtering PCR workstations

Содержание Captair Bio 320 Smart

Страница 1: ...Safer Simpler Captair Bio Smart User manual Mobile ductless filtering PCR workstations ...

Страница 2: ...control module Description of the alarms eGuard App 7 Connectivity principle Options of connection Accessing the Administrator interface Administrator interface description Fan setpoints 15 Replacing the filters 16 Recommendations for storing and using filters 19 Replacement frequency for filtration breakthrough sensor Molecode option 19 Admissible weighs on working surfaces 20 Cleaning and mainte...

Страница 3: ...ackaging and positioned correctly Erlab recommends that the electronic anemometer is calibrated at least once a year New filters must be stored in their packaging kept in a dry location see recommendations for storing and using the filters Product registration Take full advantage of the device s connectivity to enhance your safety Get up to 10 years warranty on your connected Erlab unit Register y...

Страница 4: ...inutes and 30 seconds Stick with just the instructions for the anemometer 3 Once the pulsing light stops the anemometer is calibrated and the unit is ready to be turned on 4 To operate push the power button located on the front control panel Good practices Avoid any air turbulence in the area during calibration Anemometer should be calibrated every year and every time a modification is done on the...

Страница 5: ... the front door 5 A new cycle of decontamination can be done when making sure that the front door is closed and pressing the UV button Warning UV lamp only turns on when front door is closed AND the fan ventilation is off At the end of a decontamination cycle it is recommended to turn on the fan before opening the front door UV lamp automatically switches off if the front door is opened during dec...

Страница 6: ...rotation speed RPM is 700 RPM Filtration 4 beeps 5 seconds apart Pulses Filter breakthrough Molecode S A F option The Molecode de tection value is the sensitivity setting for a period of 40s Press Mute key Please contact Erlab or your usual maintenance contact Replace filter The filter s has have reached the end of their service life lives Filter breakthrough sensor replacement Molecode 5 beeps 5 ...

Страница 7: ...net network and or local area network LAN for eGuard PC only Database Monitoring the product Monitoring controlling the product eGuard App Mobile PC Ethernet TCP IP 3 methods of connection to the Smart Portal Embedded service Mobile app PC app Conditions of use Direct connection on PC with data cable RJ45 Web connection via 3G 4G Web and or local connection Hardware requirements 1 PC 1 cable 1 App...

Страница 8: ...he settings of the unit In order to connect Use a computer equipped with an Ethernet port to plug the RJ45 cable WIFI of the computer must be switched off Web browser Internet Explorer Edge Chrome Mozilla Firefox Safari must be installed on the computer Note RJ45 cable used to plug the unit to the computer is provided 1 Direct connection on computer Check that main switch of the device is ON Take ...

Страница 9: ...tatus page and you can have access to the Settings using the following credentials Login erlab Password smart Please go to page 12 Page is not accessible Computer network parameters are not allowing the access to the embedded software Apply the following procedure Page is not accessible 3 Modify computer network parameters Right click Left click Network parameters ...

Страница 10: ...bile ductless filtering PCR workstations 10 4 Access to the Network and sharing center Left click 5 Access to the network connection Right click Left click to access properties Note This confirms your WIFI is disconnected ...

Страница 11: ...twork parameters as indicated below Left click Left click Write down your existing parameters before changing them in order to be able to set your initial parameters after the operation Enter the following parameters Validate ...

Страница 12: ...on Enter the unit IP address into the address bar of your browser in order to access the embedded sotware 1 6 7 2 3 4 5 8 9 10 11 Status page details 1 Choose active interface page 2 Device ID Model 3 Device ID serial number MAC address device status 4 Unit location 5 Molecode Option Gauge indicates the saturation level of the main carbon filter s 6 Fan Gauge indicates the fan status 7 Device use ...

Страница 13: ...13 1 5 2 3 4 6 Access to the settings is protected by the following credentials Login erlab Mot de passe smart 7 8 ...

Страница 14: ...sion of information from the device to the eGuard server for remote monitoring via eGuard App mobile PC receiving usage reports 4 Device fan setpoint settings 5 Alarm saturation filter Molecode option Sensor type indication S Solvants Sensor sensitivity settings 5 settings sensor S High sensitivity Medium High Sensitivity Medium Sensitivity Medium Low Sensitivity Low Sensitivity Sensor replacement...

Страница 15: ...1 2 Log page details 1 Displays the device s event log 2 Used for downloading the log in csv format Fan setpoints per filtration column type Unit Type of filtration column 1P 1C1P 321 2200 RPM 2600 RPM 391 ...

Страница 16: ...rkstation is used for applications other than those anticipated when the device is first set up If so please contact us so that we can verify that the current configuration is safe or it needs to be reconfigured The table below summarises all possible Flex filter technology configurations for your device Captair 321 391 481 Smart models Molecular filter HEPA filter H14 ULPA Pre filter Column Confi...

Страница 17: ...e the protective shields on either side of the workstation Diconnect the power supply to the fan Air Flow Models Captair Bio 321 391 Smart Unstack the filtration column s above the workstation After carrefuly removing the filters from their packaging assemble the column per the configuration below If your column configuration changed Apply the correct fan setpoint for the filtration column configu...

Страница 18: ...shelf life is reached we recommend to not install filters We recommend replacing the filter annually if used 24 7 Prefilters s should be replaced every year HEPA ULPA filters should be replaced every 3 years on average Molecular filter should be replaced every 2 years on average Replacement frequency of filtration breakthrough sensor Molecode option Molecular filters saturation detector Molecode S...

Страница 19: ...rylic Parts Ensure the transparency of the panels is a part of regular maintenance for the enclosure Cleaning the enclosure The enclosure must be cleaned on a regular basis This may be accomplished in several ways With soapy water followed by rinsing with clear water and drying with a soft non abrasive absorbent paper towel Or with a commercial pH neutral neutralizing product followed by drying wi...

Страница 20: ...e lower than 1 of the TLV 5 Application Questionnaire Valiquest Our laboratory specialists will recommend the appropriate filtration fume hood type of filter and personalized advice 6 Certificate of Validation for the chemicals used in the hood A certified PhD chemist issues a Certificate of Validation with a list of the chemicals approved for use in the hood 7 Our Safety Program We back up our pr...