background image

1

UVC-PKCART-ULTRA Surface 

Disinfection Unit

QUICK START GUIDE

Models:

UVC-PKCART-ULTRA

Содержание UVC-PKCART-ULTRA

Страница 1: ...1 UVC PKCART ULTRA Surface Disinfection Unit QUICK START GUIDE Models UVC PKCART ULTRA ...

Страница 2: ... such goods FOB point of original shipment by Seller at the risk and cost of Buyer Goods replaced by Seller shall become the property of Seller There is no guarantee or warranty as to anything made or sold by Seller or any services performed except as to title and freedom from encumbrances and except as herein expressly stated and particularly and without limiting the foregoing there is no guarant...

Страница 3: ...Guidelines Page 5 System Description Page 6 System Delivery Page 7 Operation Overview Page 7 Operation Settings Default Page 8 Operation Settings Customizable Page 8 Cycle Stop Status Screens Page 9 System Care Page 10 FAQ Pages 11 12 Appendix A Dosage Guidelines Page 14 INDEX ...

Страница 4: ...from the system center Contents This manual will serve as your reference guide for installation operation and service of your UV C Perma Kleen Cart PK CART Reference to all parts warranty and support are included This manual covers the system equipment products listed below Ericson Part UVC PK CART ULTRA Safety Observation It is required for the person s responsible for the installation of this eq...

Страница 5: ...r exposure to UV C however we will review safety factors regarding UV C for better knowledge and understanding Protective Clothing and Eyewear It is not recommended that any personnel be subject to direct UV C exposure In the event such exposure is probable personnel should wear PPE providing full coverage of exposed skin and appropriate eye protection UV C safety glasses designed to filter UV C a...

Страница 6: ...ides the optimum in UV dosage output to footprint served Rugged design build Ease of use controls User safety protection INDUSTRIAL STRENGTH FRAME AND STRUCTURE CORROSION PROOF CONSTRUCTION 6 INDUSTRIAL CASTERS HIGH OUTPUT PRE HEAT LAMPS FOR LONG RUN TIME LENGTH PIR SENSOR DETECTION SAFETY SHUTOFF INDIVIDUAL LAMP FAILURE INDICATOR LED S EMERGENCY STOP BUTTON BUILT TO IP56 WATER AND DUST STANDARDS ...

Страница 7: ...unit in unoccupied space centralized for operation 2 Ensure you are wearing recommended safety equipment 3 Plug system into power supply 4 Press Bacteria or Spores Operation Mode 5 Initiate system cycle 6 Evacuate space Audio and Visual Indicators 1 There is no audible alarm for countdown or run cycle on the PK CART ULTRA as not to interfere with medical facility monitoring equipment alert systems...

Страница 8: ...d on this page Target Pathogen Selection Options 1 From Start Cycle home screen user may select either Bacteria or Spores Operation Mode System default is Bacteria Mode 2 Selection may also be made by selecting User Settings and choosing Operation Mode Countdown Timer Customizing 1 From Start Cycle home screen press User Settings 2 The User Settings screen is displayed with Countdown Timer Adjustm...

Страница 9: ...ing screen and can be observed automatically adjusting with movement of the machine and selection of target Operation mode 4 Press Home button to return to Start Cycle screen for initiating treatment Cycle stop displays for normal system operation are covered on this page Emergency Stop Should cycle deactivation occur by engagement of the Emergency Stop button the following method will be required...

Страница 10: ...acement The User Setting screen will differentiate need to perform 1 Lamp Cleaning a Press Service button on home screen b Lamp Clean indicators will be illuminated in Yellow c Wipe lamps with isopropyl alcohol and clean cloth d Press Factory Settings on bottom right of screen e Enter Passcode 1470 and press Enter f Press Reset Service Light button to clear alert g Press Home to return to Start Cy...

Страница 11: ...performance with Dosimeter Cards or ATP Meter Bi Annual 1 Inspect the system interior and exterior lamp for clean conditions 2 Clean as needed 3 It is recommended to test system performance with Dosimeter Cards or ATP Meter Service Machine Message 1 Every 8 running hours it is recommended that lamps be wiped down with isopropyl or rubbing alcohol This will remove any debris that may have accumulat...

Страница 12: ...reen at the end of the cycle A Use dosimeter cards in the space to verify UVC dosage Q The touchscreen has presented a flashing Service Machine icon What does this mean A Every 8 running hours it is recommended that lamps be wiped down with isopropyl or rubbing alcohol This will remove any debris that may have accumulated on the lamp Once cleaned you can reset this message by pressing USER SETTING...

Страница 13: ...on from UVC rays Q Can I use an ATP Meter with the PK CART A Yes When used by properly trained personnel it can be used to measure how clean a surface is Q Can I use a response card with the PK CART A Yes it can be used to indicate the correct dose has been applied to a surface Q What is the difference between bacteria and sporicidal modes A Your PK CART Ultra is equipped with two modes of operati...

Страница 14: ... Spores Time in Minutes to Kill Bacteria Viruses Time in Minutes to Kill Spores 1 1 2 1 3 2 4 2 1 2 1 3 2 4 3 2 3 3 5 4 7 4 8 x 8 64 2 3 3 5 4 7 5 10 x 10 100 2 4 3 5 4 7 6 12 x 12 144 3 6 4 8 5 11 7 14 x 14 196 3 6 4 8 5 11 8 16 x 16 256 4 8 5 11 7 15 9 18 x 18 324 4 8 5 11 7 15 10 20 x 20 400 5 10 7 13 9 18 11 22 x 22 484 6 12 8 12 11 22 12 24 x 24 576 6 12 8 16 11 22 The above listed treatment ...