13 of 15
888-83-080-G-00 rev. B • 01/19
SERVICE AND MAINTENANCE
The Power Module shall be installed and repaired by qualifi ed personnel only.
The cleaning and preventive inspection should be performed by individuals with previous training.
Prior to cleaning or storing the Power Module for extended periods, the system should be placed in shut down mode. To do this,
unplug the AC power source, remove battery, then power off the Power Module by holding the Display button for 3 seconds.
Cleaning methods include:
Use a vacuum to clean out the dust from the opening of the fan and contact pin of the power bay.
Dampen cloth with water or cidex solution to clean the surface.
Do not use any alcohol or ammonia based solution to clean the surface.
Avoid spilling liquid into or on the Power Module.
When not using the battery for long periods, battery should be fully charged before storage and charged regularly during storage.
Keep spring pin contacts free of dust and debris.
OPERATING MODE
OPERATION DETAILS
AC supply mode
By connecting to AC power source, the Power Module will use the AC power to support both charging
and output loading.
The swappable battery pack and the backup battery will be charged to full when AC power is connected.
The charging current depends upon the output loading. The remaining power will be used to charge the
battery.
Battery supply mode
The Power Module will work as a battery to provide continuous AC output to varying loads.
The swappable battery pack will also charge the backup battery on the Power Module to make sure
backup battery capacity is enough to handle the swapping.
During the swapping, the display will show a 2 minute countdown and after 2 minutes, power will no
longer be provided to devices. In the event that battery fully depletes before AC power is supplied or
battery pack is replaced with a charged battery pack, the Power Module will shut down and power will no
longer be provided to devices.
Standby mode
Power Module will enter standby mode when button on the display is pressed for more than 3 seconds.
No charging activity during this mode.
AC output / Inverter shuts down.
Power Module enters low power mode.
To return back to operating mode, simply press the display button.
Shut down mode
For Power Module to enter shut down mode, both battery and AC power source must be removed.
To force the shut down mode, unplug the AC power source, remove the battery, then power off the
Power Module by holding the Display button for 3 seconds.
Alternatively, the Power Module will automatically enter shut down mode if the Battery is removed and
no new battery is placed on the Power Module. The Display will show a 2 minute count down, after
which power will no longer be provided to devices. The Display screen will then shut off and the system
will shut down.
To return back to operating mode, place a battery on the Power Module or connect AC power source to
the Power Module and press the display button.
EQUIPMENT & ACCESSORIES DISPOSAL
1. Please dispose of or recycle all batteries in accordance with local law
2. All Electronics should be recycled through an electronics recycler.
3. Remaining plastics and metals can be recycled through a commercial recycler.