![ergoline ergoselect 400 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/ergoline/ergoselect-400/ergoselect-400_operators-manual_2422184053.webp)
- 53 -
r
eCommended
seParatIon
dIstanCes
Between
PortaBle
and
moBIle
rf
CommunICatIons
equIPment
and
the
erGoseleCt
erGometer
The ergoselect ergometer is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the ergoselect ergometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the ergoselect ergometer as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum
Output Power of
Transmitter [W]
Separation Distance According to Frequency of Transmitter [m]
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
0 01
0 12
0 12
0 23
0 1
0 37
0 37
0 74
1
1 17
1 17
2 33
10
3 7
3 7
7 37
100
11 7
11 7
23 3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.