Dimensions, Weight
length:
130 cm
width:
87 cm
height:
135 cm
weight:
approx. 73 kg
Safety Standards
DIN EN 60601‑1, DIN EN 60601‑1‑2,
DIN VDE 0750‑238
Protection Class / Degree of Protection
II
/ B (arm ergometer)
BF (blood pressure module)
MDD Classification
class IIa to 93/42 EEC
RF Emission
class B to DIN EN 55011 / 5.0
DIN EN 60601‑1‑2
Environment
operation:
temperature:
+10 to +40 °C (50 to 104 °F)
rel. humidity:
30 to 75%, no condensation
atmospheric pressure: 800 to 1060 hPa
transport and storage:
temperature:
‑20 to +70 °C (‑40 to +158 °F)
rel. humidity:
10 to 95%, no condensation
atmospheric pressure: 500 to 1060 hPa
B
lood
P
ressure
m
odule
Measuring Method:
auscultatory method, oscillometric; for resting BP,
the results from both measurements are compared
for plausibility
Measuring Range
systolic pressure:
40 to 280 mmHg
diastolic pressure: 40 to 280 mmHg
pulse rate:
35 to 230 P/min
Measurement Error, systematic
systole:
+/‑ 3 mmHg
diastole:
+/‑ 3 mmHg
(temperature: +15 ... +25 °C)
Standard deviation (clinical trial)
systole / diastole: 7 mmHg (max.)
Inflation Pressure
300 mmHg max.; during inflation the inflation pressure
automatically adapts to patient's BP
Inflation Rate
between approx. 6 seconds (to 140 mmHg) and
approx. 18 seconds (to 300 mmHg)
Max Cuff Pressure
300 mmHg
Cuff Deflation Method
pulse‑dependent deflation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
Calibration
calibration with external pressure meter
Artifact Rejection
automatic artifact rejection and comparison of the rest‑
ing BP readings from both methods for plausibility