6.5
Flowchart for complete Ergometry Process
!
This ergometric process flow description is only valid for use of the ergom-
eter under internal control. When control is exercised by an ECG or PC
refer to the documentation for those devices.
Legend:
1
Start of Ergometry
Press the red coloured „
START
“ key to begin the ergometric session.
#
= input via operating
personell
= Print-out
= action via
ergometer
Selection of specific ergometric program
Assign patient to patient group, and select program for this patient accordingly, or select the
automatic program 0 for patients who you cannot be classified, or select program 9 for
patients who are to undergo heart-steady-state-training.
.
Possible modification of parameters
Prior to beginning an ergometric session you may alter any parameter for this specific ergometry.
Quiescent blood pressure measurement
Immediately after starting the ergometer, the quiescent blood pressure measurement will be
taken. It is important that the patient remains absolutely calm during this measurement as the
filter parameters for the blood pressure during stress are established by this measurement.
Artifacts during this measurement may lead to false results under stress.
Print-out of the quiescent blood pressure measurement
The resulting blood pressure data will either be printed out on the internal printer or fed to
the ECG via the RS-232 interface and documented there.
Establishment of the initial load P
o
The load program will begin with initial load Po when
the blood pressure data has been output.
First blood pressure measurement under stress
Blood pressure measurement will begin at the pre-programmed interval.
Output of load and blood pressure data
Blood pressure and load data readings are either printed by the internal printer or passed
over the RS-232 interface to an ECG for documenting there.
.
Proceeding to first load increment
The load will be incremented by the programmed first load increment as soon as data has
been output.