DISMANTLING AND ASSEMBLING
LAP BiSect scissors
Dismantling the product
The product must be cleaned and disinfected in disassembled condition with jaws opened.
1 Pull out the connecting cable
(no illustration).
2 Hold the handle (
a
) in such a way that the
moving handle is not obstructed. Press the
push-button and withdraw the shaft (
c
)
from the handle.
3 Hold the electrode by the insulation (
b
)
between your thumb and index finger.
Screw the electrode out of the shaft by
turning the rotary wheel (
e
) (approx. 6
turns).
4 Withdraw the electrode (
d
) from the shaft.
5 Pull the insulating tube (
b
) off the shaft.
Assembling the product
1 Push the insulating tube (
b
) onto the shaft
(
c
).
2 Insert the electrode (
d
) into the shaft up to
the stop position.
3 Hold the electrode by the insulation (
b
)
between your thumb and index finger.
Screw the electrode in, tightening it well,
by turning the rotary wheel (
e
) (approx. 6
turns); do not jam the thread.
4 Hold the handle (
a
) just like you would hold
a pen in such a way that the handles face
upwards and the moving handle is not ob-
structed, and press the push-button. Insert
the shaft into the handle (
a
) until it snaps
into place, without jamming it.
The complete instrument
The four disassembled components:
a Handle b Insulating tube c Shaft d Electrode