DISMANTLING AND ASSEMBLING

BiClamp

®

 LAP forceps and BiClamp

®

 E LAP forceps

Dismantling the product

The product must be cleaned and disinfected in disassembled condition with jaws opened. 

Caution: Never pull the jaws apart  

manually, as this would cause damage  

to the instrument.

1  Press both unlocking buttons on the handle 

(

d

) simultaneously and withdraw the shaft 

tube from the handle.

2   Screw the insert out of the shaft tube,  

grasping it by the joint sleeve (

e

).

3  Withdraw the insert from the shaft tube.

4   To open the jaws, hold the insert by the 

joint sleeve (

e

) and push the rod (

c

) in the 

direction of the jaws.

Assembling the product

Caution: Never press the jaws together  

manually, as this would cause damage to  

the instrument.

1   To close the jaws, hold the insert by the 

joint sleeve (

e

) and pull the rod (

b

). 

2   Insert the insert (

c

) into the shaft tube (

b

).

3  Screw in the insert up to the stop position, 

grasping it by the joint sleeve (

e

).

4   Press the rear unlocking button on the 

handle (

d

) and insert the shaft tube into the 

handle until it snaps into place at the stop 

position.

The complete instrument

The three disassembled components: 

a Handle, b Shaft tube and c Insert

d

d

e

e

b

e

c

b

d

b

e

c

a

b

For the complete notes on use, refer to 