21
propriate working order and is correctly positioned in the
groove of the cylinder connector.
WARNING!
Check that the pressure reducer is securely posi-
tioned on the argon gas cylinder. The pressure reducer
should not be loose.
8. Place the pressure line (quick coupling) onto the connection
on the rear of the APC device.
9. Connect the pressure sensor at the back of the APC device.
10. Slowly unscrew the gas cylinder valve until it is fully open.
Then turn it back again approx. a half-turn. Argon now flows
into the lines (a brief hissing sound may be heard)
WARNING!
If the hissing sound continues for more than 2
seconds, there is a leak. Close the gas cylinder valve again
immediately. Check whether the connection of the pressure
reducer and the pressure line is in fact leak-tight. The APC
device may only be used when the leak has been rectified.
5
Cleaning and disinfection
Do not sterilize!
The pressure reducer may only be cleaned from the outside. To
do so, use a cotton flannel and a medical-grade alcohol solution.
All other methods of cleaning are prohibited!
WARNING!
Flammable and explosive materials must have evap-
orated fully prior to switching on the APC device and the electro-
surgical unit.
6
Maintenance
Servicing and repair work may only be performed by competent
persons and by persons trained by ERBE. If this work is per-
formed by unqualified persons, ERBE accepts no liability and the
warranty is void.
Only use original spare parts.
OBJ_DOKU-194196-002.fm Seite 21 Freitag, 30. Oktober 2015 10:23 10
Содержание 20134-001
Страница 2: ...OBJ_DOKU 194196 002 fm Seite 2 Freitag 30 Oktober 2015 10 23 10...
Страница 65: ...65 APC 11 12 13 OBJ_DOKU 194196 002 fm Seite 65 Freitag 30 Oktober 2015 10 23 10...
Страница 66: ...66 1 2 3 4 OBJ_DOKU 194196 002 fm Seite 66 Freitag 30 Oktober 2015 10 23 10...
Страница 68: ...68 3 3 1 Erbe Elektromedizin APC Erbe OBJ_DOKU 194196 002 fm Seite 68 Freitag 30 Oktober 2015 10 23 10...
Страница 69: ...69 2 APC APC 3 2 APC 3 3 OBJ_DOKU 194196 002 fm Seite 69 Freitag 30 Oktober 2015 10 23 10...
Страница 70: ...70 APC 4 4 1 4 2 1 2 APC 3 APC 0 OBJ_DOKU 194196 002 fm Seite 70 Freitag 30 Oktober 2015 10 23 10...
Страница 71: ...71 4 APC 5 6 7 8 APC 9 APC 10 2 APC 5 OBJ_DOKU 194196 002 fm Seite 71 Freitag 30 Oktober 2015 10 23 10...
Страница 143: ...RU 20134 001 20134 002 20134 003 20134 004 20134 005 OBJ_DOKU 194196 002 fm Seite 143 Freitag 30 Oktober 2015 10 23 10...
Страница 145: ...145 APC 11 12 13 OBJ_DOKU 194196 002 fm Seite 145 Freitag 30 Oktober 2015 10 23 10...
Страница 146: ...146 1 2 3 4 OBJ_DOKU 194196 002 fm Seite 146 Freitag 30 Oktober 2015 10 23 10...
Страница 148: ...148 3 3 1 Erbe Elektromedizin APC Erbe OBJ_DOKU 194196 002 fm Seite 148 Freitag 30 Oktober 2015 10 23 10...
Страница 149: ...149 2 APC APC 3 2 APC 3 3 OBJ_DOKU 194196 002 fm Seite 149 Freitag 30 Oktober 2015 10 23 10...
Страница 150: ...150 APC 4 4 1 4 2 1 2 APC 3 APC 0 4 APC 5 OBJ_DOKU 194196 002 fm Seite 150 Freitag 30 Oktober 2015 10 23 10...
Страница 151: ...151 6 7 8 APC 9 APC 10 2 APC 5 APC OBJ_DOKU 194196 002 fm Seite 151 Freitag 30 Oktober 2015 10 23 10...
Страница 165: ...165 APC 11 12 13 OBJ_DOKU 194196 002 fm Seite 165 Freitag 30 Oktober 2015 10 23 10...
Страница 166: ...166 1 2 3 4 OBJ_DOKU 194196 002 fm Seite 166 Freitag 30 Oktober 2015 10 23 10...
Страница 168: ...168 3 3 1 Erbe Elektromedizin Erbe APC 2 APC APC 3 2 OBJ_DOKU 194196 002 fm Seite 168 Freitag 30 Oktober 2015 10 23 10...
Страница 169: ...169 APC 3 3 APC 4 4 1 4 2 1 2 APC 3 APC 0 OBJ_DOKU 194196 002 fm Seite 169 Freitag 30 Oktober 2015 10 23 10...
Страница 170: ...170 4 APC 5 6 7 O 8 APC 9 APC 10 2 APC 5 APC OBJ_DOKU 194196 002 fm Seite 170 Freitag 30 Oktober 2015 10 23 10...
Страница 172: ...172 OBJ_DOKU 194196 002 fm Seite 172 Freitag 30 Oktober 2015 10 23 10...
Страница 173: ...OBJ_DOKU 194196 002 fm Seite 173 Freitag 30 Oktober 2015 10 23 10...