Data Log Function (Option)
36
•
sending the command
<PCx>
from one of the other interfaces
•
using the display to configure the interface
Interface Parameters
DATA LOG FUNCTION (OPTION)
The device has a data log function. A memory card may be used as data logger. All measurement values will be saved,
separated by tabulators, in a text file. The time interval may be adjusted from 1-4294967 s (= 71 minutes). To activate
the data log function, a memory card must be inserted. The root directory of the memory card must contain a text file
named “EPSLOG.txt“. The new data will be written into this file.
The memory card must b e ins erted o r remo ved in st an dby
mode only!
In case, that the first entry of the first line of the file is "interval=xxxx" (xxxx = time in seconds), the memory interval is
adjusted accordingly. The entry must be written in lower case letters and without space characters. If the interval
remains unspecified, the memory interval is 60 seconds.
Example:
interval=30
The data log function is active whenever the device is not operating in standby mode. The function is indicated by a
small memory card symbol in the upper right corner of the main display. Whenever a new data set is written, the
symbol will be displayed as filled for ca. 1 second. In case the memory card is full, the symbol will be crossed out.
F
ORMAT OF THE SAVED
D
ATA
The first entry shows the present operation mode. The second entry shows the present operation mode followed by
‚U
dc
‘ and ‚I
dc
‘.
Example:
USER
I-Limit 1,0
10,02
USER
OVP 0,0
0,00
UI
U-Limit 100,01
0,10
UIP
U-Limit 100,0
0,10
SCRIPT MODE
Operation sequences may be programmed in a script which can be read in from a memory card. A script is a text file,
which includes a sequence of commands. Alternatively, the script memory can be programmed via digital interface
using the command
SCR
. For more information about the use of this command see
Commands
. The device is able
to process up to 250 commands.
E
XECUTING
/L
OADING A
S
CRIPT
The script must be saved on a MMC or SD card as text file with .txt or .scr ending.
The mode ‚Scr‘ must be selected from the main display. Pushing the rotary pulse encoder will open the file selection
menu. The script file can be selected. An error message appears if, the file could not be read correctly or, if the read in