PSI 9000 2U Series
Setting
P. Description
Repeat master init.
10 Touch area “
Initialize
” will repeat the initialisation of the master-slave system
in case the automatic enumeration of the slave units by the master is once
unsuccessful so the system would have less total power than expected or has
to be repeated manually in case the master unit couldn’t detect a missing slave
3.4.3.2
Menu “User Events”
See
„3.6.2.1 User defined events“ on page 62
3.4.3.3
Menu “Profiles”
See
„3.9 Loading and saving a user profile“ on page 64
.
3.4.3.4
Menu “Overview”
This menu page displays an overview of the set values (U, I, P or U, I, P, R) and alarm settings as well as settings
limits. These can only be displayed, not changed.
3.4.3.5
Menu “About HW, SW...”
This menu page displays an overview of device relevant data such as serial number, article number etc., as well as
an alarm history which lists the number of device alarms that probably occurred since the device has been powered.
3.4.3.6
Menu “Function Generator”
See
„3.10 The function generator“ on page 65.
3.4.3.7
Menu “Communication”
This submenu offers settings for digital communication via the optional or built-in interface. The touch area for the
installed interface module or the optional GPIB port opens one or more settings pages, depending on the interface
in use. There is furthermore an adjustable communication timeout, to make it possible to successfully transfer
fragmented messages (data packets) using higher values. In the screen for “Com Protocols” you can enable both
or disable one of the two supported communication protocols, ModBus and SCPI. This can help to avoid mixing
both protocols and to receive unreadable messages, for example when expecting a SCPI response and getting a
ModBus response instead.
For all Ethernet interfaces with two ports: „P1“ is related to port 1 and „P2“ to port 2, like printed
on the module face. Two-port interfaces will use one IP only.
IF Level 1
Description
Profibus DP
Node Address
Selection of the Profibus or node address of the device within range 1...125 via direct input
Function Tag
String input box for a user-definable text which describes the Profibus slave function tag.
Max. length: 32 characters
Location Tag
String input box for a user-definable text which describes the Profibus slave location tag.
Max. length: 22 characters
Installation Date
String input box for a user-definable text which describes the Profibus slave installation
date tag. Max. length: 40 characters
Description
String input box for a user-definable text which describes the Profibus slave.
Max. length: 54 characters
IF Level 1
Description
RS232
-
The baud rate is selectable, other serial settings can’t be changed and are defined like
this: 8 data bits, 1 stop bit, parity = none
Baud rates: 2400, 4800, 9600, 19200, 38400, 57600, 115200
IF Level 1
Description
GPIB
Node Address
Adjustment of the GPIB node address (only with option 3W installed) in the range 1...30