Message
What it means?
What to do?
Insufficient Volume error
The volume of blood or
control solution is NOT
enough.
Repeat the test with a new strip,
puncture site,
and
enough
blood/control solution.
Refer to pages
25-27
for information on
sample application.
Test result is higher than
600 mg/dL.
Re-check your glucose level.
If the glucose result is HI again, obtain
and follow instructions from your
healthcare professional without delay.
The test result is lower
than 20 mg/dL.
This may require immediate treatment
according to your healthcare
professional’s recommendations.
Although this message could be due to a
test error, it is safer to treat first and then
do another test.