both
electrodes
are of the same
type.
8a Electrodes are not
properly removed
by the jaws of the
extraction
tool
during the first run
of the device.
One or both caps are not
correctly positioned in the
extraction tool.
Check the position of the cap.
Both caps must be aligned
exactly collinear with the axis
of rotation to the extraction
tool. Even with a slight offset, it
can also lead to excessive wear
on the electrode arms.
8b Electrodes are not
properly removed
by the jaws of the
extraction
tool
during the first run
of
the
device,
wherein
wear
marks can be seen
on the electrode
holder.
One or both caps are not
correctly positioned in the
extraction tool, with the cap
retracting too far into the
extraction tool, causing the
electrode holder (sleeve or
adapter) to come in contact
with the extraction tool.
Check the position of the caps.
Both caps must be aligned
exactly collinear with the axis
of rotation to the extraction
tool. Do not drive one or both
caps too far into the extraction
tool.
9
Electrodes cannot
be driven into the
extraction
tool
because the jaws
do not open.
Wrong extraction tool with too
small
opening
for
the
electrode.
Check extraction tool and if
necessary, replace with correct
size.
Jaws do not move due to
damage
or
excessive
contamination.
Check extraction tool. Clean or
replace as needed.
Extraction
tool
is
not
approached with reversing
direction of rotation. Such a
Check program sequence and
adjust if necessary.