12 Cleaning and Maintenance
12.1 Cleaning and disinfection of device surface
Danger of electric shock
Always disconnect the device from your PC
before cleaning it.
•
Switch off the PhysioQuant Recorder.
•
Use only a moist cloth and ensure that no fluid enters the device.
The most commonly used cleaning and disinfection agents in
practices and clinics can be used.
Damage to the device
Do NOT use any disinfection agents that
contain phenol and peroxide for disinfecting
the surface of the device!
If any fluid has entered the device, then it
should only be used again after having been
checked and approved by the Service
Department.
12.2 Cleaning and disinfection of cuffs
•
Light stains can be removed by using a moist cloth.
•
For heavier stains, rinse the cuff with soapsuds or disinfected
cleaning agent (not in the washing machine!). No fluid should enter
the cuff inflation section or the connecting tube (therefore please
remove the inflation section when cleaning the cuff).
•
After you have cleaned the cuff, you need to rinse it thoroughly
with water and let it dry at room temperature for approx. 15 hours.
•
You can also use any of the following to disinfect the device:
isopropyl alcohol 70%, ethanol 70%, Microzid; Burazon liquid,
Sporicidin or Cidex. After you have disinfected the cuff, rinse it
thoroughly with water and leave it to dry at room temperature.
CAUTION!
CAUTION!