31 / 48
Max. pump pressure
Limits the max. cuff pressure to the value set between 200 mmHg and
280 mmHg.
The number of all programmed measurements will be displayed for each
measurement interval and a summary is provided for each 24 hours.
The
[Standard]
button will reset the parameters to the following values:
2 measurement intervals
Interval 1 (day phase):
06:00 - 21:59
Exact
every 15 min.
250 mmHg
Interval 2 (night phase):
22:00 - 05:59
Exact
every 30 min.
250 mmHg
After saving the selected configuration, the starting sequence of the
recorder will commence by clicking on the
[Start]
button.
The recorder will first check whether there are any previous
measurements that have not yet been read. If this is the case, then a
warning will be displayed.
If this is not the case, then the recorder will be switched off and the new
patient data and measurement intervals are stored in the recorder.
NB:
Only when a new long-term measurement is started, will the recorder
delete all old patient data and measurement values.