Doc. No. OM-101
Version 1.9: Jan. 16, 2009
16 (40)
CONTROL UNIT OPERATION (Type 2)
CONTROL UNIT OPERATION (Type 2)
Mounting Batteries
1) Open the door to the control unit.
2) Place 16 "D-cell" alkaline batteries in the holder.
Voltage
1.5V
IEC / JIS
LR 20
ANSI / NEDA
13A
3) Ensure proper battery polarity (Figure 9).
Start the control unit
The Control unit automatically starts when batteries have been installed.
Setting Temperature
1) Press the "C°/F°" button to switch between Celsius or Fahrenheit.
2) Press the set temperature button ("+/-") to enter set temperature mode.
Note!
The Set temperature mode is activated for 10 seconds. If the display returns to default mode
while adjusting the temperature, the new set temperature was not saved.
3) Press the "plus" or "minus" buttons to change the set temperature.
4) Press the set temperature button ("+/-") to confirm and return to default mode.
5) Check that the set temperature is correct on the display.
Figure 9
+
-
+
-
-
+
-
+
5.5°C
*
5.0
11.2V
11.3V
Control unit
display
Set temperature
button "+/-"
"minus"
button
"°C/°F"
button
"plus"
button
Figure 10